This document is the user manual for the temperature sensor product, which introduces the characteristics of the product, including temperature measurement, ambient temperature, Exi versions, etc.
This document serves as the installation instructions and user manual for Lapp Automaatio EPIC SENSORS multipoint temperature sensors, specifically types T-MP, T-MPT (thermocouple) and W-MP, W-MPT (resistance). The sensors are mineral insulated (MI) devices designed for professional multipoint measuring applications, available with or without enclosures and compatible with various cable materials and temperature ranges. Installation requires professional capability and adherence to safety gear and mechanical needs.
| Measuring range (Pt100) | -200…+550 °C |
|---|---|
| Measuring range (TC) | -200…+1200 °C |
| Max flange temperature | +550 °C |
| Max flange temperature (temporary) | +600 °C |
| Sensor element material | AISI 316L or INCONEL 600 |
| Flange material | AISI 316L |
| Neck pipe material | 1.4404 |
| Max measuring current (Pt100) | 1 mA |
| Max measuring current (Pt500) | 0.5 mA |
| Max measuring current (Pt1000) | 0.3 mA |
Yes, individual sensors can be delivered with own weights, or all measuring points can be covered with one common armor duct and weight.
The minimum bending radius is 2 times the outer diameter (2x ØOD) of the element.
No, grounded TCs are not allowed for Ex i certified sensor types.
The recommended maximum value is normally 1 mA to avoid false measurements or destruction of the resistor.