Engler Sonus II Ultrasonic Dental Scaler Instruction Manual

This is the instruction manual for Sonus II Ultrasonic Dental Scaler


Engler Sonus II Ultrasonic Dental Scaler Instruction Manual - cover page
Brand
Engler Engineering Corporation
Document type
User Manual
Language
English
Pages
22
File format
PDF
File size
950 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
A0161A9BC4D3AC06B14F10F2B2233D3D

About this manual

This manual provides detailed instructions for installing, operating, and maintaining the Engler Sonus II ultrasonic dental scaler. It includes safety warnings, water flow requirements, stack and tip handling, and maintenance procedures. The device uses integrated computer technology and Time Remote Feedback Circuitry to deliver ultrasonic vibrations at 18,000 Hz for effective scaling. Proper water pressure and regular maintenance are critical to avoid damage and ensure optimal performance.

The manual also covers water filter cleaning, saddle valve assembly, water tank care, and technical specifications including dimensions. Safety precautions emphasize avoiding contact with soft tissue, preventing stack bending, and ensuring adequate water flow to prevent overheating and damage. The device is intended for professional use only and is not designed to teach dentistry.

What's inside

  • READ BEFORE YOU START
  • COMPANY PROFILE
  • INTRODUCTION
  • INSTALLATION INSTRUCTIONS
  • SCALER OPERATING INSTRUCTIONS
  • SCALER MAINTENANCE
  • WATER FILTER CLEANING INSTRUCTIONS
  • SADDLE VALVE ASSEMBLY
  • WATER TANK INSTRUCTIONS
  • WATER TANK CARE & MAINTENANCE

Specifications

Operating Frequency18,000 Hz
Vibration Amplitude0.001 to 0.004 in. (25.4 um to 102 um)
Water Pressure Requirement30 to 60 PSI (0 to 4.2 kg/cm2)

Frequently asked questions

What is the operating frequency of the Sonus II?

The circuitry converts ordinary line voltage to an operating frequency of approximately 18,000 Hz.

What water pressure is required for the Sonus II?

The device must be connected to a water supply capable of delivering 30 to 60 PSI (0 to 4.2 kg/cm2).

How often should the ultrasonic stack be replaced?

The ultrasonic stack normally lasts six months to a year and should be replaced every six months to a year or as needed.