Instructions for installing, operating and maintaining the Sonus II Ultrasonic Dental Scaler
This manual provides detailed instructions for installing, operating, and maintaining the Engler Sonus II ultrasonic dental scaler. It includes safety warnings, water flow requirements, stack and tip handling, and maintenance procedures for water filters and tanks. The device uses integrated computer technology and Time Remote Feedback Circuitry to deliver ultrasonic vibrations for periodontal treatment. It is designed for professional use only and requires proper water pressure and cooling to function safely.
The manual also covers water line connections, pressure regulation, and warranty information. It emphasizes the importance of proper water flow to prevent overheating and damage to components. Users are advised to follow all safety precautions, including avoiding contact with soft tissue or skin during operation, and to replace tips and stacks regularly to maintain performance.
| Operating Frequency | 18,000 Hz |
|---|---|
| Vibration Amplitude | 0.001 to 0.004 in. (25.4 um to 102 um) |
| Water Pressure Requirement | 30 to 60 psi (2.0 to 4.2 kg/cm2) |
The internal circuitry converts ordinary line voltage to an operating frequency of approximately 18,000 Hz.
The device must be connected to a water supply capable of delivering 30 to 60 psi (2.0 to 4.2 kg/cm2).
The ultrasonic stack normally lasts six months to a year and should be replaced every six months to a year or as needed.