This manual introduces the design and use methods of ML3 soil moisture sensor, including the sensor's characteristics, maintenance, installation, configuration, calibration, troubleshooting and so on
This user manual provides instructions for installing and operating the ML3 ThetaProbe soil moisture sensor. It details cable connections, calibration procedures, and compatibility with various data loggers like GP1, GP2, DL6, and HH2. The sensor measures volumetric water content and temperature with ±1% and ±0.5°C accuracy respectively. It is designed for burial in soil and features a waterproof IP68 connector and stainless steel rods for durability. The manual also includes technical reference data, calibration methods, and troubleshooting tips.
| Volumetric Water Content Accuracy | ±1% vol over 0 to 50% vol and 0-40°C using soil specific calibrations |
|---|---|
| Temperature Accuracy | ±0.5°C over 0-40°C |
| Output Signal | 0-1V differential ≈ 0 to 60% vol nominal |
| Power Requirement | 5-14VDC, 18mA for 0.5 to 1s |
| Operating Temperature Range | -20 to +60°C |
| Environmental Rating | IP68 |
| Sample Volume | >95% influence within 40mm dia. cylinder 60mm long (approx. 75 cm3 ) around central rod. |
| Dimensions | 170.5 mm x 39.8 mm dia./138 gm (without cable) |
The sensor measures volumetric water content with an accuracy of ±1% over 0 to 50% vol and 0-40°C using soil-specific calibrations.
The sensor measures temperature with an accuracy of ±0.5°C over 0-40°C.
The ML3 is compatible with GP1, GP2, DL6, DL2e, and HH2 data loggers.
The maximum cable length for temperature measurements with DL2e is 25m.