Eplus EE820 CO2 Sensor User Guide

EE820 is a high performance CO2 sensor from Epluse. It is designed for demanding applications and supports M12 connector and RS485 communication.


Eplus EE820 CO2 Sensor User Guide - cover page
Brand
E-Plus
Category
sensor
Document type
User Manual
Language
English
Pages
7
File format
PDF
File size
3.1 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
95E43B716161DB85B767BAF528969FE3

About this manual

This document serves as the user guide for the Eplus EE820 CO2 sensor, designed for demanding applications. It details the scope of supply, including mounting sets and accessories, and provides critical safety cautions regarding mechanical stress and environmental protection. The guide explains the sensor's dimensions for conduit installation and describes the connection diagrams for both analogue and RS485 output versions.

What's inside

  • Scope of Supply
  • Accessories and Spare Parts
  • Dimensions for Conduit Installation
  • Connection Diagrams
  • Technical Data
  • Setup and Adjustment
  • Digital Interface Hardware
  • Modbus Setup
  • Maintenance
  • Replacement and Retrofit

Specifications

Measuring principledual wavelength non-dispersive infrared technology (NDIR)
Measurement range0...2 000 / 5 000 / 10 000 ppm
Response time t63300 s (standard) / 140 s (fast)
Supply voltage24 V AC ±20% / 15 - 35 V DC
Working conditions temperature-20...60 °C
Protection classIP54

Frequently asked questions

Does the EE820 require user configuration?

The EE820 is ready to use and does not require any configuration by the user, as the factory setup corresponds to the ordered type number.

What is the factory setting for the Modbus slave address?

The factory setting for the Modbus slave address (EE820) is 67.

Can the output signal be changed from voltage to current?

Yes, the output signal can be changed by setting the output signal selection switch to I for current or U for voltage.

Download PDF (3.1 MB)