E+E Elektronik EE31 Series Humidity/Temperature Transmitter Manual

This document is the Hardware and Software Manual of E+E Elektronik® Ges.m.b.H.'s EE31 series humidity/temperature transmitter.


E+E Elektronik EE31 Series Humidity/Temperature Transmitter Manual - cover page
Brand
E+E Elektronik
Document type
User Manual
Language
English
Pages
40
File format
PDF
File size
4.7 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
F9852FBABE2E3BF85DA01EC00B3D19FD

About this manual

This manual provides detailed instructions for installing, calibrating, and maintaining E+E Elektronik's EE31 series humidity and temperature transmitters. It covers various mounting models (A, B, D, H, E) and their specific applications, including wall, duct, and pressure-tight installations. The document also includes safety guidelines, electrical connections, and troubleshooting procedures for optimal operation.

The manual is intended for qualified personnel and includes technical specifications, configuration software guidance, and maintenance steps such as sensor and probe replacement. It emphasizes safety precautions, especially when handling the sensitive sensor elements and during installation of pressure-tight probes.

What's inside

  • General
  • Product Description
  • Installation
  • Electrical Connections
  • Operating Components
  • Alarm Module (optional)
  • Calibration
  • Maintenance
  • Network
  • Scope of Supply

Specifications

Working Range (Sensor Probe, Model D)-40...180°C (-40...356°F)
Working Range (Electronics)-40...60°C (-40...140°F)
Voltage Range8...35V DC; 12...30V AC (option C03/C07)

Frequently asked questions

What are the different mounting models for the EE31 series?

The EE31 series includes models A (wall mounting), B (duct mounting), D and H (remote sensing probe), E (pressure-tight sensor), and F (wall mounting with rear cable outlet).

What are the working temperature ranges for the EE31 series models?

The working range for most models is -40...60°C (-40...140°F), while Model B and Model H sensors can operate up to 80°C (176°F) and Model D sensors up to 180°C (356°F).

What is the maximum pressure rating for the Model E sensor probe?

The Model E sensor probe is suitable for applications at pressures between 0.01...20bar (0.15...300psi).

How should the sensor probe be oriented when installing the transmitter?

The sensor probe must be mounted pointing downwards, either horizontally or vertically.