Installation and operating manual for sample gas pumps P1.1, P1.1E
The document serves as an official user guide for the Bühler Technologies P1.1 and P1.1E measurement gas pumps. It details the scope of delivery, product descriptions, and specific installation requirements, including mounting on vibration dampers and handling high moisture samples by inverting the pump head.
Key operational sections cover starting and controlling the pump, as well as comprehensive maintenance procedures such as replacing check valves, O-rings, bellows, and crank gears. The manual also addresses safety warnings regarding hazardous gases, electrical voltage, and hot surfaces, along with disposal instructions. This manual is written in Chinese.
The motor's rated environmental temperature allows installation at altitudes up to 1000 m.
The manual suggests using vibration pads with a diameter of 10 mm, a height of 10 mm, and Shore hardness 70.
A spacing of at least 20 mm must be maintained between the motor and the back wall.