This manual introduces the use of Super Systems DPC2530 Continuous Digital Dew Point Analyzer. The analyzer can be used to measure endothermic atmosphere, exothermic atmosphere, nitrogen / hydrogen atmosphere and plant air systems dew point. The manual also introduces the instrument's specifications, power supply, display type, digital communications, alarms, etc.
The Super Systems DPC2530 is a continuous digital dew point analyzer designed for measuring endothermic, exothermic, nitrogen/hydrogen atmospheres, and plant air systems. The instrument features an LED digital display with a measurement range of -0 to +80°F (-47 to +27°C) and supports RS485 Modbus communication as well as 4-20 mA retransmission output.
The manual provides detailed instructions on terminal block wiring for power, communications, and alarm outputs. It outlines the startup procedure, which includes allowing the unit to stabilize at ambient temperature before operation to prevent sensor condensation. Operational guidance covers connecting sampling lines, adjusting flow rates between 1.0 and 2.0 CFH, and interpreting display readings.
| Measurement Range | -0 to +80°F (-47 to +27°C) |
|---|---|
| Temperature Range | 0 to 120°F (-18 to +49°C) |
| Power Supply | 115 VAC 60Hz (Factory set to 115 or 240 VAC) |
| Display Type | LED Digital Display |
| Resolution | +/- 1°F (+/- 0.1°C) |
| Digital Communications | RS485 Modbus |
| Control Output | 4 –20 mA |
| Alarms | Two Alarm Relays |
| Size (Closed Case) | 11”H x 10”W x 7”D |
| Weight | Approximately 8 lb. |
The analyzer is suitable for Endothermic Atmosphere, Exothermic Atmosphere, Nitrogen / Hydrogen Atmosphere, and Plant Air Systems.
The optimum flow rate should be between 1.5 and 2.0 cubic feet per hour (CFH), though a rate as low as 1.0 CFH is acceptable.
Moisture cannot be allowed to enter; if the sensor becomes wet, use an inert gas like nitrogen or dry compressed air to dry it.
Under no circumstances should methane or natural gas be used to dry the DPC2530.