This manual covers the operation, setup, calibration, and configuration of the Super Systems M4582 Basic Nitriding Sampling System for measuring hydrogen…
The Super Systems M4582 Basic Nitriding Sampling System is designed to measure hydrogen concentration (H2), display the percentage of hydrogen (H2), dissociation rate (DA), and ammonia (NH3). It also includes functionality to calculate nitrogen potential (Kn) based on input parameters. The system features a 3.5-inch color LCD touchscreen interface and supports various communication protocols including RS485 Modbus and Ethernet.
| Power Supply | 100-240 VAC |
|---|---|
| Current | Max. 0.2 Amps |
| Sensor Principle | Thermal Conductivity |
| User Interface | 3.5" Color LCD Touchscreen |
| Measurement Range | 0-100% H2 |
| Hydrogen Measurement Accuracy | +/- 0.01% |
| Hydrogen Measurement Repeatability | +/- 0.01% |
| Hydrogen Measurement Resolution | 0.01% |
| Analog Output | 2 Independent 4-20mA |
| Analog Output Resolution | 0.005mA |
| Analog Output Accuracy | +/- 0.01% |
| Analog Output Linearity | +/- 0.01% |
| Analog Output Load Resistance | Min 0 Ohms, Max 500 Ohms |
| Digital Communication | RS485 Modbus, Ethernet |
| Dimensions | Approx 13" L x 13" W x 6" H |
| Weight | 22.5 lbs / 10.2 kg |
| Recommended Flow Rate | 1.5 to 2.0 SCFH |
| Operating Environment Humidity | 10-90 %RH (Non-condensing) |
| Operating Temperature | 32 to 122°F / 0 to 50°C |
| Sample Gas Temperature | 32 to 158°F / 0 to 70°C |
The device can display Percent Hydrogen (H2), Percent Dissociation (DA), Percent Ammonia (NH3), and Nitriding Potential (Kn).
Kn is calculated using the formula Kn = pNH3/pH2 3/2, requiring input of N2, NH3, and DA values.
The recommended flow rate is between 1.5 and 2.0 SCFH (0.71 to 0.94 lpm).
Sensor calibration requires two gases: pure nitrogen or argon for zero calibration, and a span gas with hydrogen content close to the process gas.