The DP2000 Portable Dew Point Analyzer Operations Manual is provided by Super Systems Inc. It introduces the specifications, warnings, startup, operation, and maintenance of the DP2000 Portable Dew Point Analyzer.
This document serves as the operations manual for the Super Systems DP2000 Portable Digital Dew Point Analyzer. It provides comprehensive instructions for the startup, operation, maintenance, calibration, and troubleshooting of the instrument. The manual details how to connect sampling tubes, adjust flow rates, and interpret readings for various applications such as heat treating furnaces and endothermic generators.
| Dew Point Range | 0 to 80°F (-18 to 27°C) |
|---|---|
| Temperature Range | 0 to 120°F (-18 to 49°C) |
| Power Supply | 115 or 240 VAC |
| Display Type | LED Digital Display |
| Resolution | +/- 1°F (+/- 0.1°C) |
| Battery Life | Approx. 8 hours |
| Size (Closed Case) | 11" x 10" x 7" |
| Weight | 8.2lbs (3.7 kg) |
The optimum flow rate of the sample gas should be between 1.5 and 2.0 cubic feet per hour (CFH), although a flow rate as low as 1.0 CFH is acceptable.
Regardless of the application, allow the DP2000 to sample gas for two to three minutes before recording any data to ensure the sensor has achieved equilibrium.
The two main causes of damage to the DP2000 are the ingestion of soot and water, which cause erroneous readings and long-term damage to the sensor.