This manual describes how to operate the FlowPrep ™ Sample Preparation System.
The document serves as an operator guide for the FlowPrep degasser, a laboratory instrument designed to remove contaminants from samples using heat and inert gas. It details the safety precautions, intended use, and general specifications of the device.
The manual provides comprehensive instructions on installing the system, including unpacking, connecting gases, and attaching delivery tubes. It further explains sample preparation procedures such as controlling heating station temperatures, cleaning sample tubes, and determining sample mass for various tube types.
| Sample Tube OD Capacity | Up to 3.0 cm (1.18 in.) |
|---|---|
| Temperature Range | 10 ºC above ambient to 400 ºC |
| Temperature Accuracy | within ± 10 °C |
| Heating Block Voltage | 100/120/ 230/240 VAC ± 10% |
| Maximum Power | 250 VA maximum |
| Frequency | 50/60 Hz |
| Operating Temperature | 10 to 35 ºC |
| Gases Used | Helium, nitrogen, argon |
| Maximum Flow Rate | Up to 50 cm3/min. |
| Weight | 8.6 kg (19 lbs) |
The instrument uses Helium, nitrogen, argon, or any other pure gas that does not adsorb at room temperature.
The system can accommodate sample tubes with an outer diameter of up to 3.0 cm (1.18 in.).
The instrument operates within an electrical temperature range of 10 to 35 ºC.