This manual introduces how to operate FlowPrep 060 and the matters need to be noticed during operation.
The FlowPrep 060 Degasser prepares samples for adsorption analysis by using flowing gas passed over a heated sample to remove moisture and other contaminants. The instrument features six heating stations and six cooling stations, with the degassing temperature easily set on the front temperature controller pad. Gas flow control valves provide a constant indication of the valve state.
The document covers general information, installation procedures including unpacking and power selection, operation and sample preparation for various tube types, and troubleshooting and maintenance tasks such as checking gas flow and cleaning delivery tubes.
| Heating Block Temperature Range | 10 ºC above ambient to 400 ºC |
|---|---|
| Electrical Voltage | 100, 120, 220, 240 V AC ± 10% |
| Maximum Power | 250 V A |
| Frequency | 50/60 Hz |
| Operating Temperature | 10 to 35 ºC |
| Non-operating Temperature | 0 to 50 ºC |
| Humidity | 45 to 80% relative, non-condensing |
| Gases Used | Helium, nitrogen, argon, or any other pure gas that does not adsorb at room temperature |
| Maximum Flow Rate | Up to 50 cm3/min |
| Vacuum Requirement | No vacuum required |
| Sample Capacity | 6 heating stations and 6 cooling stations |
| Maximum Sample Tube OD | Up to 3.0-cm (1.18-in.) |
| Physical Height | 45 cm (16.93 in.) |
| Physical Width | 36 cm (14.17 in.) |
| Physical Depth | 30 cm (11.81 in.) |
| Weight | 8.6 kg (19 lbs) |
The instrument uses Helium, nitrogen, argon, or any other pure gas that does not adsorb at room temperature.
No, the FlowPrep 060 does not require a vacuum to operate.
The heating block temperature can be set up to 400 ºC.
The unit has six heating stations and six cooling stations.