Official preinstallation guide for the Micromeritics TriStar II Plus analyzer, covering site preparation, unpacking, space requirements, power specs, gas…
This document provides instructions and a checklist for preparing a laboratory site to install the Micromeritics TriStar II Plus analyzer. It covers essential steps such as unpacking and inspecting shipping damage, verifying instrument space dimensions, and ensuring environmental factors like temperature and humidity are within specified limits.
The guide details specific requirements for power supply (100/115/230 V AC), computer system compatibility (Windows 7 Professional or higher), and gas supply hardware. It emphasizes the use of standard copper gas lines and compatible regulators to ensure accurate measurements, while also outlining safety precautions and a commitment statement for installation.
| Operating Temperature | 15 to 35 °C |
|---|---|
| Non-operating Temperature | 0 to 50 °C |
| Relative Humidity | 20 to 80% relative, non-condensing |
| Power Input Voltage | 100 / 115 / 230 V AC |
| Power Frequency | 50 or 60 Hz |
| Instrument Height | 74 cm (29 in.) |
| Minimum Gas Pressure | 200 psig (1378.9 kPag) |
Standard installation requires 1/8 in. copper gas supply lines located in the instrument accessories kit.
Gas bottles must be placed within 6 feet (1.8 m) of the instrument's rear or left side.