This is a service manual for a remote blood oxygen saturation monitoring system
The Nellcor Intouch Remote Oximetry Notification System is an alarm notification system designed for real-time monitoring of routine patient status and alarm events from specific Nellcor pulse oximetry monitors. It serves as a parallel, redundant mechanism to inform clinical staff of patient alarms in hospital environments. The system consists of three basic components: the Intouch Transmitter, which connects to the monitor; the Intouch Pager, worn by caregivers; and the Intouch Programmer, used for setup.
| Single Alarm Response Time | 10 seconds |
|---|---|
| Multiple Alarm Response Time | 60 seconds |
| Maximum Transmitters per Frequency | 24 |
| Input Voltage (Transmitter) | +15 volts, 1.5 amps |
Only qualified service personnel should service this product.
Patient alarms will be received by the pager within 10 seconds for single patient alarm events.
Up to 24 transmitters and 12 pagers may operate on one frequency and be programmable by one programmer.
The transmitter requires an input voltage of +15 volts at 1.5 amps.