Nellcor N-180 Pulse Oximeter Service Manual

This document is the maintenance manual of N-180 pulse oximeter


Nellcor N-180 Pulse Oximeter Service Manual - cover page
Brand
Nellcor
Document type
Service Manual
Language
English
Pages
25
File format
PDF
File size
472 KB
Published
16 October, 2023
Updated
16 October, 2023
Model
N-180
MD5 checksum
FE5EB342A99244E0C835F56AEAD4CA14

About this manual

The document serves as a technical guide for servicing the Nellcor N-180 Pulse Oximeter. It details performance and safety tests, including patient source current and sink current leakage measurements using an electrical safety analyzer.

Key specifications are outlined, covering environmental requirements such as operating temperature ranges (50-104°F) and altitude limits (up to 10,000 ft). The manual also lists electrical characteristics like voltage compatibility (100-240 VAC), power consumption, and battery details. A comprehensive spare parts list is provided at the end of the document.

What's inside

  • Performance and Safety Tests
  • Specifications
  • Spare Parts Listings
  • Environmental Requirements
  • Electrical Characteristics
  • Physical Characteristics

Specifications

Saturation Range0-100%
Pulse Rate Range20-250 bpm
Operating Temperature10-40°C
Storage Temperature0-50°C
Maximum Altitude10,000 ft
Voltage Input100-240 VAC
Battery TypeLead-acid battery pack, 1.9 AH

Frequently asked questions

What is the maximum leakage current for a 220 V monitor?

The analyzer leakage indication must not exceed 50 microamps.

What is the minimum recharge period for the battery?

The minimum recharge period is 14 hours.

Download PDF (472 KB)