This document is the user manual for the Genius TM 3 Tympanic Thermometer and Base. The Genius 3 Tympanic Thermometer is an ear canal thermometer with measurement site adjusted modes, which include oral and rectal temperatures.
The Genius™ 3 Tympanic Thermometer is an adjusted mode ear thermometer designed to provide fast and accurate measurements of patient temperatures from the tympanic membrane. It also offers equivalent measurement modes for oral and rectal temperatures based on the tympanic reading. The device features a Peak Select System, temperature recall, and various display icons and alarms to assist operators.
| Temperature measurement range | 33.0°C to 42.0°C (91.4°F to 107.6°F) |
|---|---|
| Measurement time | 1 - 2 seconds |
| Operating Humidity limitation | 15% to 90% |
| Storage and transportation temperature range | -25°C to 55°C (-13°F to 131°F) |
The thermometer provides measurements from the tympanic membrane, as well as equivalent oral and rectal temperatures.
The device provides temperature measurement in 1 - 2 seconds.
No, do not use the thermometer on patients with ear drainage, blood, brain or spinal fluid, vernix, ear wax plugs, or foreign bodies in the ear canal.
Allow the thermometer to acclimate to room temperature for a minimum of 30 minutes prior to use.