This manual introduces calibration methods for the 700G series pressure gauge.
The document serves as a comprehensive guide for calibrating the Fluke 700G Series Pressure Gauges. It details how to enter calibration mode, perform verification tests using required equipment, and adjust settings via serial interface commands.
Safety information is provided regarding electrical shock prevention, battery handling, and environmental limits. The manual also lists available input ranges, accuracy specifications, media compatibility, and mechanical dimensions for various models within the series.
| Accuracy (Positive Pressure) | ±0.05 % FS |
|---|---|
| Accuracy (Vacuum) | ±0.1 % FS |
| Operating Temperature | -10 °C to +55 °C |
| Battery Life (Continuous) | 1,500 hours |
The warranty period is three years and begins on the date of shipment.
If the display reads "OL", the range limit is exceeded and the pressure source must be removed immediately.
The maximum allowed torque is 13.5 Nm (10 ft lb).