Fluke 754 HART Mode Users Guide

The 754 HART Mode Users Guide introduces the features and characteristics of 754 HART Mode, including HART Channel Selection, Connect to a HART Transmitter, etc.


Fluke 754 HART Mode Users Guide - cover page
Brand
Fluke
Document type
User Manual
Language
English
Pages
22
File format
PDF
File size
523 KB
Published
15 October, 2023
Updated
15 October, 2023
MD5 checksum
351625EC0289234ADB8D672817C7044A

About this manual

The Fluke 754 is a tool that combines communication and calibration functions in one unit. This specific guide details the procedures for using the device's HART mode to communicate with HART transmitters via its serial interface or mA jack.

Users can connect to supported or generic transmitters, view active device information such as model numbers and range values, and perform various operations including reading process variables and duplicating basic setup data. The guide also covers service operations like Loop Test, Output Trim, and Sensor Trim for adjusting transmitter blocks.

What's inside

  • HART Channel Selection
  • Connect to a HART Transmitter
  • Supported vs. Generic Transmitters
  • See Process Variables
  • Basic Setup Operations
  • Sensor Calibration
  • Device Identification
  • Loop Test
  • Output Trim
  • Duplicate Basic Information

Specifications

Internal Series Resistance for Loop Power250 Ω
External Loop Supply Resistance Rangebetween 230 Ω and 270 Ω

Frequently asked questions

Can I perform a Sensor Trim on any transmitter?

No, you can only do a Sensor Trim on a supported transmitter.

What internal resistance does the Fluke 754 use when supplying loop power?

The product supplies loop power through an internal series resistance of 250 Ω.

What external resistance is required if using an external loop power supply?

A resistance between 230 Ω and 270 Ω must be connected in series with the external supply.

Which HART operations are available for generic transmitters?

Generic transmitters allow access to the Active Device screen, Basic settings, Sensor data (read only), Device Identification, HART Output, Setup, HART Information, Pressure Zero Trim, and Process details.

Download PDF (523 KB)

More Fluke Manuals


View all Fluke manuals →