This manual introduces the use of Fluke 1587 and 1577 insulation multimeters, including function introduction, usage steps, maintenance and maintenance, etc.
This manual provides comprehensive operating instructions for the Fluke 1587 and 1577 insulation multimeters. It details safety precautions, including warnings about unsafe voltage and test lead alerts, and explains the functions of the rotary switch and buttons. The guide covers basic measurements such as AC/DC voltage, resistance, capacitance, frequency, and temperature, as well as specific insulation testing procedures.
| Maximum Voltage Applied to any Terminal | 1000 V ac rms or dc |
|---|---|
| Operating Temperature | -20 °C to 55 °C |
| Battery Type | Four AA batteries |
| Weight | 550 g |
| IP Rating | IP40 |
| Insulation Measurement Range (Model 1587) | 0.01 M Ω to 2 G Ω |
| Insulation Measurement Range (Model 1577) | 0.1 M Ω to 600 M Ω |
| Test Voltages (Model 1587) | 50, 100, 250, 500, 1000 V |
| Test Voltages (Model 1577) | 500, 1000 V |
| Short-Circuit Test Current | 1 mA nominal |
The warranty period is three years and begins on the date of shipment.
The meter requires four AA batteries (NEDA 15A or IEC LR6).
The maximum voltage applied to any terminal is 1000 V ac rms or dc.
The discharge time is less than 0.5 second for a load of 1 μF or less.