This document is a user manual for FX0241107 tabletop centrifuges, including usage methods and precautions
This is the Kubota instruction and service manual, document FX0241107, for the Model 2420 table-top centrifuge (October 2005). It explains the component names and control panel, installation and power supply, grounding, and operation: opening and closing the lid, setting speed by rpm or by centrifugal force (×g), setting the timer, acceleration/deceleration, saving memory, and function settings such as rotation radius, SLOW deceleration, end-of-operation sound, reminder alarm and indicators.
The service sections cover daily, monthly and annual inspections, cleaning and sterilization of the chamber, rotor, buckets and tube rack, greasing, inspection of the circuit protector and use of the photoelectric tachometer port. Further sections show how to mount the RS-240, RS-1004 and RMP-23 swinging bucket rotors, list alarm and error indicators for troubleshooting, and give specifications, standard accessories and recommended spare parts. The text notes the products are intended for operators with expert knowledge and states the manual is not to be distributed in the U.S.A., Mexico, Canada and Australia.
| Intended use | Preprocessing in in-vitro analysis; not designed to connect directly to a patient's body |
|---|---|
| Materials that must not be separated | Hazardous material (explosive, chemically active, organic, radiation containing, or contaminated by pathogenic microorganisms) or oil |
| Warranty period | One year after delivery or fifteen months after shipping, whichever comes earlier |
It can be used for preprocessing in in-vitro analysis, breaking down blood or urine samples containing plasmas or cells. It is not designed to connect directly to a patient's body.
Do not use it for hazardous material such as explosive, chemically active, organic or radiation containing material, material contaminated by pathogenic microorganisms, or oil.
Kubota repairs or replaces a defective instrument free of charge if the defect is proven within one year after delivery or fifteen months after shipping, whichever comes earlier.
The manual covers mounting of the RS-240, RS-1004 and RMP-23 rotors, including their swinging bucket versions.