This document provides installation precautions for ELARA11, including unpacking inspection, mechanical inspection, and installation preparations.
This guide provides step-by-step instructions for installing the Tuttnauer Elara11 autoclave. It covers unpacking procedures, mechanical inspection for damage, and requirements for the installation surface, including load capacity and ventilation space.
The document details necessary utility connections, specifically power supply specifications (230Vac), grounding requirements, and safety checks like leakage current tests. It concludes with initial startup steps, including running a Vacuum Test and B&D Test to verify proper operation before daily use.
The counter must be at least 22 inches wide and 24 inches deep.
You must keep the back and sides of the unit 2 inches away from the wall.
The room ventilation must be at least 10 cycles per hour.