This document provides installation precautions for ELARA11, including unpacking inspection, mechanical inspection, and installation preparations.


SYSTEM GENERAL SG6841 handbook - cover page
Brand
SYSTEM
Category
medical technology
Document type
Installation Guide
Language
English
Pages
3
File format
PDF
File size
877 KB
Published
07 August, 2012
Updated
29 September, 2023
MD5 checksum
04BAE43B10AEDD62729165B41535E1F8

About this manual

This guide provides step-by-step instructions for installing the Tuttnauer Elara11 autoclave. It covers unpacking procedures, mechanical inspection for damage, and requirements for the installation surface, including load capacity and ventilation space.

The document details necessary utility connections, specifically power supply specifications (230Vac), grounding requirements, and safety checks like leakage current tests. It concludes with initial startup steps, including running a Vacuum Test and B&D Test to verify proper operation before daily use.

What's inside

  • Unpacking the autoclave
  • Installation preparations
  • Connections to utility supplies
  • Initial operational checks
  • Vacuum Test procedure
  • B&D Test cycle

Frequently asked questions

What counter dimensions are required for the Elara11 autoclave?

The counter must be at least 22 inches wide and 24 inches deep.

How much space must be left between the autoclave and the wall?

You must keep the back and sides of the unit 2 inches away from the wall.

What is the minimum required room ventilation rate?

The room ventilation must be at least 10 cycles per hour.