Kangaroo Entriflush Feeding Pump Directions For Use

This document provides the directions for use of KANGAROO® ENTRIFLUSH® feeding pump, including normal operation and battery operation. It also explains the steps to insert the Kendall Healthcare Kangaroo Entriflush Pump Administration Sets into the Entriflush Pump.


Kangaroo Entriflush Feeding Pump Directions For Use - cover page
Brand
Kangaroo
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
86 KB
Published
03 August, 2012
Updated
28 September, 2023
MD5 checksum
0BA69257ED35131686E34E8255D98892

About this manual

This document provides directions for use for the Kangaroo Entriflush Feeding Pump. It details the setup process, including how to insert the Kendall Healthcare Kangaroo Entriflush Pump Administration Sets, and explains how to operate the device using either AC power or battery power. The guide covers setting hourly feeding rates, configuring flush doses and intervals, and utilizing the 'Flush Now' feature.

What's inside

  • NORMAL OPERATION
  • BATTERY OPERATION
  • FEED DOSE
  • FLUSH DOSE
  • FLUSH NOW FEATURE
  • No Set Alarm
  • Feed Error Alarm
  • Dose Delivered Alarm
  • Bad Set Alarm
  • Flush Error Alarm

Frequently asked questions

How do I recharge the pump's battery?

To recharge the battery, plug the pump into an AC wall outlet. The pump will automatically begin charging.

What causes the 'Feed Error Alarm'?

The alarm is triggered if the feeding container is empty, the feeding tube or set is occluded, or the drip chamber walls are coated with formula.

What should I do if the 'Bad Set Alarm' occurs?

Check for a fluid leak in the set valve by inserting the valve and pressing 'RUN HOLD'. If the set has been in use for 24 hours, replace it.

Download PDF (86 KB)