CADD Prizm VIP 6100/6101 Ambulatory Infusion Pump Operator's Manual

This manual concerns only the CADD-Prizm® VIP (Variable Infusion Profile) Model 6100 and Model 6101 ambulatory infusion pumps.


CADD Prizm VIP 6100/6101 Ambulatory Infusion Pump Operator's Manual - cover page
Brand
CADD
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
126
File format
PDF
File size
1.4 MB
Published
03 August, 2012
Updated
30 September, 2023
MD5 checksum
CFB30F40E2A24FB9B8E3BD9B954D33B1

About this manual

This Operator's Manual is intended for clinician use only and covers the CADD-Prizm VIP (Variable Infusion Profile) Model 6100 and Model 6101 ambulatory infusion pumps. The document details the pump's delivery modes, including PCA, Continuous, TPN, and Intermittent, and provides instructions for programming and operation.

What's inside

  • General Description and Basic Operations
  • Pump Setup and Programming
  • PCA Delivery Method
  • Security Codes
  • Reference & Troubleshooting
  • Warnings
  • Cautions

Specifications

System Delivery Accuracy+6%

Frequently asked questions

Who is this manual intended for?

This manual is intended for clinician use only and should not be accessed by patients.

What are the operating temperature limits for the pump?

Do not operate the pump at temperatures below +2°C (36°F) or above 40°C (104°F).

What is the maximum bolus volume that can be set?

The amount of the bolus can be set as high as 20 ml (or the mg or mcg equivalent).

What happens if the Air Detector is not installed or turned off?

The pump will not detect air in the fluid path, so you must periodically inspect the fluid path to prevent air embolism.