CADD Prizm PCS/Pain Control System 6100/6001 OPERATOR'S MANUAL

This document is the operator's manual for the CADD-Prizm® PCS (Pain Control System) Model 6100 and 6101 ambulatory infusion pumps. The pump can be programmed to deliver medication at a constant rate and/or allow delivery of a bolus dose at a specified time interval. This manual is intended for clinician use only and should not be accessed by patients.


CADD Prizm PCS/Pain Control System 6100/6001 OPERATOR'S MANUAL - cover page
Brand
CADD
Category
Other products
Document type
User Manual
Language
English
Pages
102
File format
PDF
File size
1.4 MB
Published
03 August, 2012
Updated
29 September, 2023
MD5 checksum
81B09B150291512AFA3DCDBAB06C1FC6

About this manual

This document is the Operator's Manual for the CADD-Prizm® PCS (Pain Control System) Model 6100 and 6101 ambulatory infusion pumps. It is intended strictly for clinician use to program the device for delivering medication at a constant rate or allowing bolus doses at specified intervals.

What's inside

  • Warnings
  • Cautions
  • General Description
  • Pump Setup and Programming
  • Operating the Pump
  • Stopping the Pump
  • Starting a Clinician Bolus
  • Starting a Demand Dose

Specifications

System Delivery Accuracy+6%
Operating Temperature Range+2°C to 40°C (36°F to 104°F)
Storage Temperature Range-20°C to 60°C (-4°F to 140°F)
Storage Humidity Range10% to 90% relative humidity

Frequently asked questions

Who is this manual intended for?

This manual is intended for clinician use only and should not be accessed by patients.

Can the pump deliver blood or cellular blood products?

No, the pump is not designed for the delivery of blood or cellular blood products.

What type of batteries should not be used with this pump?

Do not use rechargeable NiCad, nickel metal hydride (NiMH), or carbon zinc ('heavy duty') batteries.

Download PDF (1.4 MB)