CN-0271 is a K-type thermocouple measurement system composed of AD8495, AD8476, AD7790, ADR441 with integrated cold junction compensation.
The document details a complete thermocouple signal conditioning circuit designed for measuring K-type thermocouple temperatures. The system utilizes the AD8495 amplifier for cold junction compensation and signal amplification, driving an AD7790 16-bit sigma-delta ADC via an AD8476 differential amplifier. The circuit provides a compact, low-cost solution for high-resolution analog-to-digital conversion.
| Temperature Range | 15°C to 950°C |
|---|---|
| Output Voltage Calibration | 5 mV/°C |
| ADC Resolution | 16-bit |
| Noise Free Resolution | 13.4 bits |
| Linearity Error | within 2°C |
| System Error with Correction | within 1°C |
The system is calibrated for a temperature range of 15°C to 950°C.
The AD8495 thermocouple amplifier provides integrated cold junction compensation.
The measured noise free resolution is approximately 13.4 bits.