Hanna Instruments HI 98140/HI 98150 Portable Waterproof pH/ORP Meters Instruction Manual

HI 98140 and HI 98150 are two of the most innovative portable pH meters ever offered. These instruments feature IP 67 waterproof protection, GLP (Good Laboratory Practice) capabilities and use the HANNA instruments®’ intelligent electrodes. These pH electrodes incorporate a chip which memorizes the calibration data performed with a specific instrument. When the “intelligent” electrode is attached to the meter again, it is automatically recognized. This technology allows the operator to optimize time and efficiency with unsurpassed safety. It avoids erroneous measurements taken in the event the pH electrode is substituted after calibration.


Hanna Instruments HI 98140/HI 98150 Portable Waterproof pH/ORP Meters Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
17
File format
PDF
File size
995 KB
Published
10 July, 2012
Updated
28 September, 2023
MD5 checksum
3FAED0B14429D052DF935917EEA7B5BB

About this manual

This document is the instruction manual for the Hanna Instruments HI 98140 and HI 98150 Portable Waterproof pH/ORP Meters. These instruments feature IP 67 waterproof protection, GLP capabilities, and utilize intelligent electrodes that memorize calibration data to ensure measurement safety and efficiency.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Initial Preparation
  • Setup Mode
  • Taking pH Measurements
  • Taking ORP Measurements
  • Calibration Procedure
  • Logging Function

Specifications

Range (pH)-4.00 to 19.99 pH
Resolution (pH)0.01 pH
Accuracy (pH @20°C/68°F)+/-0.01 pH
Waterproof ProtectionIP 67
Power Supply4x 1.5V AA batteries / approx. 300 hours of use (with backlight off) or 12 Vdc input
Dimensions196 x 80 x 60 mm (7.7 x3.1x 2.4)
Weight500 g (1.1 Ib.)

Frequently asked questions

What is the factory setting for the password?

The factory setting for the password is 0000.

How long does the meter automatically switch off after non-use?

The meter is automatically switched off after 5 minutes of non-use.