Hanna Instruments HI 93705 Silica ISM Instruction Manual

This manual introduces the product features, usage methods and other information of Hanna HI 93705 Silica ISM


Hanna Instruments HI 93705 Silica ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
327 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
2F7E233720FF806BF2D6C57BB1C64A72

About this manual

The HI 93705 is a portable instrument designed to measure silica (SiO2) content in water and wastewater within a range of 0.00 to 2.00 mg/L. It utilizes an exclusive positive-locking system for cuvet placement and features liquid and powder reagents dosed precisely for maximum repeatability.

What's inside

  • Warranty
  • Preliminary Examination
  • General Description
  • Specifications
  • Required Reagents
  • Measurement Procedure
  • Interferences
  • Tips for Accurate Measurement
  • Battery Replacement

Specifications

Range0.00 to 2.00 mg/L
Resolution0.01 mg/L
Accuracy+/-0.03 mg/L +3% of reading
Light SourceLED @ 890 nm
Battery Life40 hours
Auto-Shut offAfter 10 minutes of non-use

Frequently asked questions

What is the measurement method used?

It uses the heteropoly blue method, adapted from ASTM D859.

How long does it take to replace the battery?

The manual states that replacement must only take place in a non-hazardous area using a 9V alkaline battery.

What causes interference with phosphate?

Phosphate above 60 mg/L causes a 2% reduction, and above 75 mg/L causes an 11% reduction.