This document is the instruction manual of Hanna HI3810 Dissolved Oxygen Test Kit
The Hanna Instruments HI 3810 Dissolved Oxygen Test Kit is a portable device used to determine oxygen concentration in water quickly and easily. It utilizes a modified Winkler method (Azide Modification Titration) and is suitable for use in both field and laboratory environments. The kit includes all necessary reagents and calibrated vessels for performing up to 110 tests on average.
| Range | 0 to 10 mg/L (ppm) |
|---|---|
| Smallest Increment | 0.1 mg/L (ppm) |
| Analysis Method | Azide Modification Titration |
| Sample Size | 5 ml |
| Number of Tests | 110 (average) |
The kit uses a modified Winkler method known as Azide Modification Titration.
Multiply the milliliters of titration solution read from the syringe scale by 10 to obtain mg/L (ppm) oxygen.
The smallest increment is 0.1 mg/L (ppm).