Hanna Instruments HI 3818 Carbon Dioxide Test Kit Instruction Manual

This instruction manual provides instructions for using the Hanna HI 3818 Carbon Dioxide Test Kit. The kit can be used to quantitatively measure carbon dioxide levels in water. The measurement range is 0-40 mg/L, with an accuracy of 0.5 mg/L.


Hanna Instruments HI 3818 Carbon Dioxide Test Kit Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
217 KB
Published
09 July, 2012
Updated
30 September, 2023
MD5 checksum
4B447DC2F458FA7313F34C4BB28C035F

About this manual

The Hanna Instruments HI 3818 is a portable chemical test kit designed for measuring carbon dioxide (CO2) levels in water. It includes phenolphthalein indicator, calibrated vessels of varying sizes (5 mL, 10 mL, and 50 mL), and a calibrated syringe containing the titration solution. The kit allows users to perform tests in both field and laboratory settings.

What's inside

  • Specifications
  • Significance and Use
  • Chemical Reaction
  • Determination in the 0 to 100 mg/L Carbon Dioxide
  • Determination in the 0 to 50 mg/L Carbon Dioxide
  • Determination in the 0 to 10 mg/L Carbon Dioxide
  • Accessories

Specifications

Range (0-100 mg/L)0 to 100 mg/L (ppm) CO2
Range (0-50 mg/L)0 to 50 mg/L (ppm) CO2
Range (0-10 mg/L)0 to 10 mg/L (ppm) CO2
Smallest Increment (0-10 range)0.1 mg/L
Analysis MethodBase titration using phenolphthalein indicator
Sample Size Options5 ml, 10 ml and 50 ml
Number of Tests (Average)110 tests

Frequently asked questions

What is the chemical reaction method for this test kit?

Carbon dioxide in the water sample is neutralized with a dilute sodium-hydroxide solution to a pH of 8.3 using a phenolphthalein indicator.

How do you calculate the result for the 0-10 mg/L range?

Multiply the values on the syringe scale by 10.

What happens if the solution turns pink during testing?

The titration is complete, and you should record the result based on the volume of titration solution used.

Download PDF (217 KB)