This two-page quick reference guide from Agilent Technologies (Stratagene Products) gives an optimized protocol for using the Brilliant III Ultra-Fast SYBR Green QPCR Master Mix with the StepOnePlus Real-Time PCR System from Applied Biosystems. It covers preparing the reactions, setting up the QPCR plate and thermal profile in the StepOnePlus software, running the PCR program and analyzing data. For detailed instructions the guide refers users to the full product manual.
The reaction mixture is built from 2× SYBR Green QPCR Master Mix, upstream and downstream primers at optimized concentration, diluted reference dye and nuclease-free PCR-grade water to a final volume of 20 μl, with suggested DNA quantity ranges for genomic DNA and cDNA. The guide also lists catalog numbers, ordering information, technical services contacts and notices to purchaser.
| Catalog number (400 reactions) | 600805 |
|---|---|
| Catalog number (4000 reactions) | 600816 |
| Final reaction volume | 20 μl |
| Primer concentration | 200–500 nM (optimized) |
| Reference dye dilution | 1:50 |
| Genomic DNA per reaction | 5 pg – 50 ng |
| cDNA per reaction | 0.5 pg – 100 ng |
| Diluted reference dye per reaction | 0.3 μl |
The protocol uses a final reaction volume of 20 μl, and the run method is set to 20 μl.
The suggested range is 5 pg – 50 ng for genomic DNA and 0.5 pg – 100 ng for cDNA.
Select SYBR Green Reagents (including a melt curve) and the Fast ramp speed.
Dilute the reference dye 1:50 using nuclease-free PCR-grade water.