Agilent P3211 and P3311 pH Combination Electrodes Operating Guide

The Agilent Technologies Agilent P3211 and P3311 pH Combination Electrodes are used to measure pH in aqueous solutions. The refillable reference electrode helps to prolong the service life of the electrode. They are used with the 3200P pH Meter or similar meters. The P3211 consists of a pH-sensing electrode and reference electrode. The P3311 consists of a pH-sensing electrode, reference electrode, and temperature-sensing probe.


Agilent P3211 and P3311 pH Combination Electrodes Operating Guide - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
19
File format
PDF
File size
888 KB
Published
06 June, 2012
Updated
29 September, 2023
MD5 checksum
7C5328129B099DC7A57E3B7AF8750CF0

About this manual

This document provides the operating instructions for the Agilent P3211 and P3311 Refillable pH Combination Electrodes. These electrodes are designed to measure pH in aqueous solutions and are compatible with the 3200P pH Meter or similar instruments. The P3211 model consists of a pH-sensing electrode and a reference electrode, while the P3311 includes an additional temperature-sensing probe.

What's inside

  • Overview
  • Operation
  • Cleaning
  • Maintenance
  • Electrode Restoration
  • Troubleshooting

Specifications

pH range0 to 14
Zero potential7.00 ± 0.5
Temperature range (P3211)0 to 80 ºC
Temperature accuracy± 0.5 °C
Probe impedance< 300 Mohm
Reference typeAg/AgCl
Body materialGlass
Probe diameter12 mm

Frequently asked questions

Which meters are compatible with these electrodes?

They are used with the 3200P pH Meter or similar meters.

What is the recommended calibration solution for acidic samples?

For acidic sample solutions, use pH 4.00 and pH 7.00 standards.

How should protein precipitation on the electrode be cleaned?

Soak the measuring tip in a 0.1 mol/L HCl solution that contains 1% pepsin for 15 minutes.

What is the procedure to restore a damaged glass sensitive membrane?

Soak the sensitive membrane in 4% HF solution for 3 to 5 seconds, then rinse with HCl and water.