MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual

MAS-100™ is a professional microbial air monitoring system developed and produced by MBV AG in Switzerland. It is suitable for detecting microorganisms in the air.


MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual - cover page
Brand
Operator’s
Category
Other
Document type
User Manual
Language
English
Pages
34
File format
PDF
File size
605 KB
Published
08 March, 2012
Updated
30 September, 2023
MD5 checksum
93587B8D9C4852028AE519DF755E07AB

About this manual

The MAS-100™ is a high‑performance microbial air monitoring instrument that draws 100 L of air per minute and deposits airborne microorganisms onto standard Petri dishes. It follows the Andersen impaction principle and complies with ISO 14698‑1 for clean‑room monitoring. The manual explains the basic principle, safety considerations and the advantages of the device.

The guide details all operating functions: turning the unit on and off, selecting manual, template or selectable sampling modes, adjusting volumes (1‑2000 L), using the delay function, calibrating the instrument, and connecting a printer. It also provides maintenance tips, battery handling, and a full list of technical specifications.

What's inside

  • Introduction
  • Basic principle
  • General functions
  • Manual air sampling
  • Template mode
  • Selectable mode
  • Instrument settings
  • Calibration
  • Technical specifications
  • Accessories and spare parts

Specifications

Flow rate100 L/min ± 2.5 %
Standard sampling volumes50, 100, 250, 500, 1000 L
Freely definable sampling volume range1 – 2000 L
Sampling head diameter110 mm
Overall device weight2.2 kg
Battery pack9‑cell NiMH, 10.8 V

Frequently asked questions

How do you start a manual air sampling?

After turning the MAS‑100 on, press “yes” at the “Start?” prompt to begin the manual sampling cycle.

How can the sampling volume be changed?

Select the “Change volumes?” option in Parameter Mode and set any value between 0 and 2000 L.

What is the default calibration reminder period?

The calibration reminder is preset to 12 months by the manufacturer, but it can be set from 0 to 12 months.