MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual
The MAS-100™ is a high‑performance microbial air monitoring instrument that draws 100 L of air per minute and deposits airborne microorganisms onto standard Petri dishes. It follows the Andersen impaction principle and complies with ISO 14698‑1 for clean‑room monitoring. The manual explains the basic principle, safety considerations and the advantages of the device.
The guide details all operating functions: turning the unit on and off, selecting manual, template or selectable sampling modes, adjusting volumes (1‑2000 L), using the delay function, calibrating the instrument, and connecting a printer. It also provides maintenance tips, battery handling, and a full list of technical specifications.
Covers: Introduction · Basic principle · General functions · Manual air sampling · Template mode · Selectable mode
Latest Operator’s manuals
MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual
User Manual 34 pages · 605 KB · ENOperator's manual for the MBV MAS-100 microbial air sampler, covering setup, sampling modes, calibration and technical specifications.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in Operator’s manuals
How do you start a manual air sampling?
After turning the MAS‑100 on, press “yes” at the “Start?” prompt to begin the manual sampling cycle.
From: MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual
How can the sampling volume be changed?
Select the “Change volumes?” option in Parameter Mode and set any value between 0 and 2000 L.
From: MBV MAS-100 Professional Microbial Air Monitoring System Operator's Manual