Promega PowerPlex Matrix Standards 3100 Technical Operation Manual

This document describes the usage of PowerPlex® Matrix Standards 3100 kit, including the generation of spectral calibration files and amplification of sample DNA.


Promega PowerPlex Matrix Standards 3100 Technical Operation Manual - cover page
Brand
PowerPlex
Category
Other
Document type
User Manual
Language
Chinese
Pages
8
File format
PDF
File size
478 KB
Updated
28 September, 2023
MD5 checksum
39FB06440089F8DFD49228FAAAC82ED4

About this manual

This manual describes the Promega PowerPlex Matrix Standards 3100 kit used to generate spectral calibration files on the ABI PRISM 3100 genetic analyzer. It covers kit components, sample preparation, instrument setup, spectral calibration runs, results criteria, DNA amplification references, and analysis of amplified samples.

The kit contains four fluorescently labeled DNA fragments (JOE, FL, TMR, and CXR) and is designed for PowerPlex 16 and 1.2 systems and other GenePrint fluorescent STR systems. This manual is written in Chinese.

What's inside

  • I. 介绍
  • II. 产品组分
  • III. 在 ABI PRISM 3100 遗传分析仪上产生 PowerPlex 光谱校正文件
  • A. 光谱运行模式
  • B. 光谱参数
  • C. Matrix 样品制备
  • D. 仪器准备
  • E. 光谱校正运行
  • F. 光谱校正结果
  • IV. 样品 DNA 扩增

Specifications

Spectral calibration Q-valueGreater than 0.95

Frequently asked questions

How should the Matrix Standards be stored?

All components are stored at -20°C. The Matrix standard fragments are light-sensitive and must be kept in the dark, and the components are recommended to be stored and used together with amplified reagents.

Which systems are supported by PowerPlex Matrix Standards 3100?

The spectral calibration files are applicable to PowerPlex 16, PowerPlex 1.2, and other GenePrint fluorescent STR systems.

What are the acceptance criteria for spectral calibration?

The Q-value must be greater than 0.95 and the condition number must be in the range 6.0 to 9.0.

How many capillaries should pass calibration?

At least 12 of the 16 capillaries should pass calibration; otherwise spectral calibration must be repeated.