Siemens Echomax XRS-5 Ultrasonic Level Sensor User Manual


Siemens Echomax XRS-5 Ultrasonic Level Sensor User Manual - cover page
Brand
Siemens
Category
Ultrasonic level meter
Document type
User Manual
Language
Chinese
Pages
10
File format
PDF
Published
17 December, 2011
Updated
29 September, 2023
Model
XRS-5
MD5 checksum
A1AB89C64F259852EB3798B72F7D56DF

About this manual

This manual outlines the basic features and functions of the Siemens Echomax XRS-5 series ultrasonic level sensors, which are used together with Siemens Milltronics ultrasonic level monitoring products. The sensor converts electrical pulses from the transceiver into acoustic energy and converts the returning acoustic energy back into electrical energy for the controller, and it has a built-in temperature sensor.

The document covers general safety guidelines including use in hazardous areas, specifications, dimensions, mounting and interconnection. Wiring sections describe direct connection with shielded twisted-pair cable, coaxial cable connection via a junction box, and twisted-pair extension connection, and the appendix gives installation drawings for the XRS-5 and XRS-5C in hazardous locations. This manual is written in Chinese.

What's inside

  • Introduction
  • General guidelines
  • Specifications
  • Dimensions
  • Mounting
  • Interconnection
  • Direct connection
  • Coaxial cable connection
  • Twisted-pair extension connection
  • Appendix A: Installation drawings

Specifications

Measurement range0.3 to 8 m
Frequency43 kHz
Ambient temperature-20 to 65 °C
Ingress protectionIP68
Weight1.2 kg
Cable typeShielded twisted pair, 0.5 mm² (18 AWG), PVC jacket
Maximum distance to transceiver365 m with RG-62A/U coaxial cable

Frequently asked questions

What is the measuring range of the XRS-5 sensor?

The measurement range is 0.3 to 8 m for liquids and slurries.

What ambient temperature can the sensor be used in?

The sensor series is approved for use only at ambient temperatures from -20 °C to 65 °C.

What is the maximum distance between the sensor and the transceiver?

With RG-62A/U coaxial cable the connection run can be up to 365 m.

How is the sensor powered?

The sensor can only be powered by a Milltronics dedicated controller.

Download PDF

More Siemens Manuals


View all Siemens manuals →