GMO Hunter 18S Real-Time PCR Operating Manual

This manual describes the method of using the Hunter 18S fluorescence real-time PCR kit produced by Haikang Life Technology Co., Ltd. to detect 18S gene sequence.


GMO Hunter 18S Real-Time PCR Operating Manual - cover page
Brand
GMO
Document type
User Manual
Language
Chinese
Pages
1
File format
PDF
File size
200 KB
Published
28 November, 2011
Updated
30 September, 2023
MD5 checksum
35EA344EB00DBA109F6FE085BBBD8D44

About this manual

The GMO Hunter 18S manual provides step‑by‑step instructions for preparing and running the 18S real‑time PCR assay. It details reagent composition, storage requirements, and the PCR cycling program required for accurate detection.

Additional sections cover control setup, data analysis guidance, warranty information, and contact details for technical support from Haikang Life Technology. This manual is written in Chinese.

What's inside

  • Reagent Kit Composition
  • Reagent Storage
  • PCR Reaction Setup
  • Cycling Conditions
  • Control Setup
  • Data Analysis
  • Technical Support
  • Warranty and Liability

Specifications

Reagent storage temperature-20 °C
Mastermix volume per reaction11.2 µL
Water volume per reaction8.8 µL (adjusted for template)
Fluorescence detection wavelength530 nm (FAM)

Frequently asked questions

How should the PCR reagents be stored?

Reagents and positive controls must be stored at –20 °C and kept protected from light, for example by wrapping tubes in aluminum foil.

How many replicates and controls are required per sample?

Each sample should be run in two PCR replicates, with a positive control (2 µL) and a negative control (2 µL water) included in every run.