Yokogawa GA10/GA10CL/GA10UP Error E3049: Failed to operate some projects or all project at once.

Yokogawa GA10/GA10CL/GA10UP · taken from the manufacturer’s manual, pages cited

Error codeE3049

Failed to operate some projects or all project at once. The manual's first action is: “Close projects that do not require data collection. To start recording, change the data file save destination folder.”

Model
Yokogawa GA10/GA10CL/GA10UP
Display shows
Failed to operate some projects or all project at once. (E3049)
Type
Error code
Source
Manual p. 154

What E3049 means

Failed to operate some projects or all project at once. p. 154

Common causes

  • If Start Monitoring Simultaneously cannot be executed, it could be any of the following reasons. The number of simultaneously running projects or simultaneously connected devices exceeds the limit. There is not enough available memory on Windows. p. 154

What to do

  1. Close projects that do not require data collection. To start recording, change the data file save destination folder. p. 154

If E3049 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Yokogawa service.

This page summarises what the manufacturer’s manual says about E3049; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Yokogawa. Found a mistake? Report an error.