Error, fault and alarm codes of products - what each code means, its causes and what to do - read from the manufacturers’ manuals, with the manual page each answer comes from.
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
An Olympus HF instrument has not been properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
The Olympus HF instrument is not properly connected to the UNIVERSAL socket. The manual's first action is: “Check the connection of the Olympus HF instrument to the UNIVERSAL socket.”
The power level for the chosen mode is set to zero. The manual's first action is: “Increase the power level via the "Set screen" (refer to chapter 5.4, "Output setting").”
The device indicates no mode has been selected while activating. The manual's first action is: “Select a mode via the "Mode screen" (refer to chapter 5.4, "Output setting").”
The double pedal footswitch has not been assigned to the corresponding output socket. The manual's first action is: “Assign the double pedal footswitch via the "Assign Footswitch screen" (refer to chapter 6.4, "Assign…
The single pedal footswitch has not been assigned to the corresponding output socket. The manual's first action is: “Assign the single pedal footswitch via the "Assign Footswitch screen" (refer to chapter 6.4, "Assign…
The maximum time limit for the application has been exceeded. The manual's first action is: “Release the footswitch or handswitch and activate again by repressing the footswitch or handswitch.”
The screen is touched while switching on or there's a malfunction of the electrosurgical generator. The manual's first action is: “Release the finger from the screen.”
The handswitch of the HF instrument is pressed while switching on. The manual's first action is: “Release the handswitch of the HF instrument.”
The handswitch of the HF instrument is pressed while switching on. The manual's first action is: “Release the handswitch of the HF instrument.”
The footswitch pedal is pressed while switching on. The manual's first action is: “Release the footswitch pedal.”
The footswitch pedal is pressed while switching on. The manual's first action is: “Release the footswitch pedal.”
The push button was pressed while switching on or there is a malfunction. The manual's first action is: “Release the push button.”
The footswitch combination error E019 indicates two single or double pedal footswitches are connected or the generator is malfunctioning. The manual's first action is: “Ensure that only one single pedal and / or only …