Mitek VAPR II / VAPR 3 Error 400: Footswitch BLUE pedal stuck

Electrosurgical device · taken from the manufacturer’s manual, pages cited

Fault code400

Error 400 means the BLUE footswitch pedal is stuck, the YELLOW footswitch pedal is stuck, or there's a handswitch INPUT fault. The manual's first action is: “Depress and release the MODE button once.”

Model
Mitek VAPR II / VAPR 3
Display shows
Footswitch BLUE pedal stuck (400)
Type
Fault code
Source
Manual p. 35

What 400 means

Footswitch BLUE pedal stuck p. 35

Common causes

  • Footswitch BLUE pedal stuck p. 35
  • Footswitch YELLOW pedal stuck p. 35
  • Hands witch INPUT fault p. 35
  • Menu/reset button stuck p. 35
  • Electrode identification circuit fault p. 35
  • Front panel switch fault: yellow UP button p. 35
  • Front panel switch fault: yellow DOWN button p. 35
  • Front panel switch fault: blue UP button p. 35
  • Front panel switch fault: blue DOWN button p. 35
  • Front panel switch fault: MODE button p. 35
  • Intermittent activation switch p. 35
  • Accessory thermistor fault (open) p. 35
  • Accessory thermistor fault (short) p. 35
  • Accessory Fault (invalid electrode configuration) p. 35

What to do

  1. Depress and release the MODE button once. p. 35

Before you service the product, read the safety instructions in the manual p. 10.

If 400 is still not resolved

  • Contact Mitek technical service for assistance. p. 10
    “Contact Mitek technical service for assistance.”

This page summarises what the manufacturer’s manual says about 400; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Mitek. Found a mistake? Report an error.