Keysight 654xA, 655xA, 664xA, 665xA Error E11: Ambient temperature read back test failed (power-on, and for models 664xA & 665xA, also *TST?).

GPIB DC Power Supply · taken from the manufacturer’s manual, pages cited

Error codeE11

Ambient temperature read back test failed (power-on, and for models 664xA & 665xA, also *TST?) due to defective thermistor AIRT770 or comparator AIU513. The manual lists no action for this code.

Model
Keysight 654xA, 655xA, 664xA, 665xA
Display shows
E11
Type
Error code
Source
Manual p. 37

What E11 means

Ambient temperature read back test failed (power-on, and for models 664xA & 665xA, also *TST?). p. 37

Common causes

  • Thermistor AIRT770 or comparator AIU513 defective. p. 37

If E11 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Keysight service.

This page summarises what the manufacturer’s manual says about E11; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Keysight. Found a mistake? Report an error.