HP 8903B Error 30: Input overload detector tripped in range hold.

HP 8903B Audio Analyzer · taken from the manufacturer’s manual, pages cited

Error code30

30 on the HP 8903B is listed in the manual as: Input overload detector tripped in range hold.. The manual's first action is: “Press CLEAR key and then enter a more realistic range setting, or press AUTOMATIC OPERATION key to allow the Audio Analyzer to seek the correct input range. Refer to Automatic Operation.”

Model
HP 8903B
Display shows
Input overload detector tripped in range hold. (30)
Type
Error code
Source
Manual p. 112

What 30 means

Input overload detector tripped in range hold. p. 112

Common causes

  • This error code indicates that the input signal is too high for the selected range. p. 112

What to do

  1. Press CLEAR key and then enter a more realistic range setting, or press AUTOMATIC OPERATION key to allow the Audio Analyzer to seek the correct input range. Refer to Automatic Operation. p. 112

Before you service the product, read the safety instructions in the manual p. 4.

If 30 is still not resolved

  • Refer servicing to qualified personnel. p. 5
    “Refer servicing to qualified personnel.”
  • Servicing instructions are for use by qualified personnel only. p. 5
    “Servicing instructions are for use by qualified personnel only.”

This page summarises what the manufacturer’s manual says about 30; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with HP. Found a mistake? Report an error.