Fluke 5700A/5720A Series II Error Codes

501 codes from pages 199–295 of the manual

This is the list of fault, error codes the Fluke manual gives for the 5700A/5720A Series II Fluke 5700A/5720A Series II. Find your code below and open it for what the manual says about its causes and what to do.

Fault codes from p. 199

CodeWhat it meansPage
CME Command error (fault) p. 199
EXE Execution error (fault) p. 199
DDE Device-dependent error (fault) p. 199
QYE Query error (fault) p. 199
100 Low Air Flow p. 273
201 5725 Self-Test ROM Failure p. 273
202 5725 Self-Test RAM Failure p. 273
203 5725 Self-Test EEPROM Failure p. 273
204 5725 Self-Test Data Bus Failure p. 273
205 5725 Self-Test CLAMPS Circuit Failure p. 273
206 5725 Self-Test HVCLR Circuit Failure p. 273
207 5725 Self-Test DAC Failure p. 273
208 5725 Self-Test Watchdog Timer Failure p. 273
209 5725 Current Heatsink Too Hot p. 273
210 Output Tripped To Standby p. 273
211 5725 Current Compliance Voltage Too High p. 273
212 5725 Current Compliance Voltage Too High p. 273
213 5725 +400V Supply Did Not Shut Off p. 273
214 5725 -400V Supply Did Not Shut Off p. 273
215 5725 Voltage Heatsink Too Hot p. 273
216 5725 Voltage Heatsink Too Hot p. 273
217 5725 +400V Supply Too Small p. 273
218 5725 +400V Supply Too Large p. 273
219 5725 -400V Supply Too Large p. 273
220 5725 -400V Supply Too Small p. 273
221 5725 +400V Supply Current Too High p. 274
222 Output Tripped To Standby p. 274
223 5725 -400V Supply Current Too High p. 274
224 Output Tripped To Standby p. 274
225 5725 Fan Not Working p. 274
226 5725 CLAMPS Fault p. 274
227 Output Tripped To Standby p. 274
228 5725 Software TRAP p. 274
229 5725 Cable Was Off p. 274
230 5725 RESET (power-up or watchdog timer) p. 274
231 5725 Guard-Crossing Timeout p. 274
232 5725 Illegal/Unexecutable Command p. 274
233 5725 Non-Maskable Interrupt Occurred p. 274
234 5725 HVCLEAR Circuit Activated p. 274
235 Output Tripped To Standby p. 274
400 Bad Cal Constant ID p. 274
500 Bad Mode To CNFmodeRanges p. 274
501 Configuration Table Overflow p. 274
502 Bad Amplifier Type Selection p. 274
503 Bad Flat Constant Type Selection p. 274
504 Bad Flat Constant Type In Table p. 274
505 5725 Guard Crossing Wouldn't Start p. 274
600 Bad Row Selector p. 274
601 Bad Column Argument p. 274
602 Bad Character p. 274
800 Wideband Module Needed p. 274
801 Amplifier Not Selected Or Connected p. 274
802 Amplifier Must Remain On For This Output p. 274
803 Over Limits Of Locked Range p. 274
804 Under Limits Of Locked Range p. 274
1104 Bad Control Byte From 5725 Receive Task p. 277
1105 Illegal 5725 Transmit Task State p. 277
1106 5725 Indefinite ACKWAIT Holdoff p. 277
1200 Serial Write Failure To 5725 p. 277
1201 Packet Too Large For 5725 p. 277
1300 Undefined Signal p. 277
1301 Output Tripped To Standby p. 277
1302 Output Tripped To Standby p. 277
1303 DC Zero Calibration is Needed (Every 30 Days) p. 277
1400 Cannot Install MTick() p. 277
1401 Bad Timer Selector p. 277
1500 Could not ACK Packet From Inguard p. 277
1501 Illegal Inguard Receive Task State p. 277
1502 Bad Receive Packet Number From Inguard p. 277
1503 Bad Control Byte From Inguard p. 277
1600 Multiple Timeouts Sending To Inguard p. 277
1601 Inguard Request Reset Loop p. 277
1602 Unexpected NSA From Inguard p. 277
1603 Bad Packet Number In ACK From Inguard p. 277
1604 Bad Control Byte From Inguard Rcv Task p. 277
1605 Illegal Inguard Transmit Task State p. 277
1606 Inguard Indefinite ACKWAIT Holdoff p. 277
1700 Serial Write Failure To Inguard p. 277
1701 Packet Too Large For Inguard p. 277
1801 Bad Transition Type p. 277
1802 Bad Boolean Selector p. 278
1803 NRMmgStby Encountered Error p. 278
1804 DCV Called For Non-DCV Range p. 278
1805 ACV Called For Non-ACV Range p. 278
1806 Bad ACV Frequency Range p. 278
1807 Hi Res Frequency Too High p. 278
1808 ACV Amplitude Correction Failure p. 278
1809 Cannot Phase Lock To Signal p. 278
1810 Bad Phase Quadrant p. 278
1811 Current Called for Non-Current Range p. 278
1812 Ohms Called for Non-Ohms Range p. 278
1813 Cannot Phase Lock, Output Not AC p. 278
1814 Cannot Phase Shift, Output Not AC p. 278
1815 Bad Wide Band Range p. 278
1816 Freq. Too High Even For Wide Band p. 278
1817 Illegal Current Output Location p. 278
1818 Output Current Out Of Tolerance p. 278
1819 Current Compliance Voltage Exceeded p. 278
1820 2-Wire Compensation Current Exceeded p. 278
1821 NRMbstcur Passed Bad Range p. 278
1822 5220 No Longer Connected p. 278
1823 5220 Detected Fault And Shut Off p. 278
1824 5205 No Longer Connected p. 278
1825 5205 Detected Fault And Shut Off p. 278
1826 5725 No Longer Connected p. 278
1827 14 Bit DAC Scaling Found Zero Output p. 278
1900 Could not Write Byte To EEPROM p. 278
1901 Checksum Error Reading EEPROM p. 278
1902 EEPROM Is Full p. 278
1903 Unknown Nonvolatile Constant Selector p. 278
1904 EEPROM Partition Is Full p. 278
1905 Could not Write Byte To 5725 EEPROM p. 278
1906 Checksum Error Reading 5725 EEPROM p. 278
2000 Bad Command Code p. 279
2001 Bad Signal p. 279
2002 Long Term Command In Progress p. 279
2003 Guard Crossing Protocol Failed To Start p. 279
2004 Fatal Fault, Output Tripped p. 279
2005 Tripped While Calibrating or Diagnosing p. 279
2100 Bad Display Selector p. 279
2101 Bad Character p. 279
2102 Bad Annunciator Selector p. 279
2200 Unknown Command p. 279
2201 Invalid Number Of Parameters p. 279
2202 Invalid Cal Constant Name p. 279
2203 Invalid Keyword p. 279
2204 Invalid Range p. 279
2205 Invalid Parameter Type p. 279
2206 Invalid Parameter Unit p. 279
2207 Invalid Parameter Value p. 279
2208 IEEE488.2 I/O DEADLOCK p. 279
2210 IEEE488.2 INTERRUPTED Query p. 279
2211 Not Allowed From GPIB Interface p. 279
2212 Not Allowed From Serial Interface p. 279
2213 Remote Only p. 279
2214 Invalid Syntax p. 279
2215 IEEE488.2 UNTERMINATED Command p. 279
2216 Symbol Table Overflow p. 279
2217 Command Reserved For Service p. 279
2218 Invalid Binary Number p. 279
2219 Invalid Binary Block p. 279
2220 Invalid Character p. 279
2579 Could not Cal 19 Ohm (Meas 1) p. 283
2580 Could not Cal 19 Ohm (Meas 2) p. 283
2581 Could not Ext Cal 1 Ohm (0.26V CM Meas) p. 283
2582 Could not Ext Cal 1 Ohm (Ext Meas) p. 283
2583 Could not Ext Cal 1 Ohm (0.13V CM Meas) p. 283
2584 Could not Ext Cal 1 Ohm (Int Meas) p. 283
2585 Could not Ext Cal 1.9 Ohm (0.18V CM Meas) p. 283
2586 Could not Ext Cal 1.9 Ohm (0.12V CM Meas) p. 283
2587 Could not Ext Cal 1.9 Ohm (Int Meas) p. 283
2588 Could not Cal Wideband Gain p. 283
2589 Bad Cal Step Selector p. 283
2590 A/D Appears To Have Failed p. 283
2591 User Range Gain Adjustment Is 0 p. 283
2592 Stored User Range Gain Adjust Is 0 p. 283
2593 220 µA AC Flatness Calibration Failed p. 283
2594 2.2 mA AC Flatness Calibration Failed p. 283
2595 22 mA AC Flatness Calibration Failed p. 283
2596 220 mA AC Flatness Calibration Failed p. 283
2597 Bad Range Number For IAC Flatness Cal p. 283
2598 No Data For DCI Calibration p. 283
2599 2.2V AC Flat Early Fail p. 283
2600 2.2V AC Flat Bad AC Cal Sensor? p. 283
2601 2.2V AC Flat Diverge p. 283
2602 22V AC Flat Early Fail p. 283
2603 22V AC Flat Bad AC Cal Sensor? p. 283
2604 22V AC Flat Diverge p. 283
2605 220V AC Flat Early Fail p. 283
2606 220V AC Flat Bad AC Cal Sensor? p. 283
2607 220V AC Flat Diverge p. 283
2608 1100V AC Flat Early Fail p. 283
2609 1100V AC Flat Bad AC Cal Sensor? p. 283
2610 1100V AC Flat Diverge p. 283
2611 220 µA IAC Flat Early Fail p. 283
2646 DAC Appears To Not Meet Linearity Spec p. 285
2647 ACV Buffer Offset Too Large p. 285
2648 5725 ACV Flat Test Cap Too Large p. 285
2649 5725 ACV Flat Test Cap Too Large p. 285
2650 Could not Cal 100M Ohm (Meas 2) p. 285
2700 A7: 8255 Control Word p. 285
2701 A7: 8255 Port A Fault p. 285
2702 A7: 8255 Port B Fault p. 285
2703 A7: 8255 Port C Fault p. 285
2705 A7: Oven Regulation Fault p. 285
2706 A7: Current Compliance Fault p. 285
2707 A7: Hardware Initialization Fault p. 285
2709 A7: Current Magnitude Fault p. 285
2710 A7: Dummy Load Current Fault p. 285
2711 Assembly A7 Is Not Responding p. 285
2712 A7: Hi-Res Clock Fault p. 285
2713 A7: Hi-Res Loop Fault In 100HZ Range p. 285
2714 A7: Hi-Res Loop Fault In 1KHZ Range p. 285
2715 A7: Hi-Res Loop Fault In 10KHZ Range p. 285
2716 A7: Hi-Res Loop Fault In 100KHZ Range p. 285
2717 A7: Hi-Res Loop Fault In 1MHZ Range p. 285
2718 Fault In Setting Up AC For Diagnostics p. 285
2719 A7: 8255 Was Reset p. 285
2800 A11: 8255 Control Word p. 285
2801 A11: 8255 Port A Wires p. 285
2802 A11: 8255 Port B Data Bus p. 285
2803 A11: 82C54 Status Words p. 285
2804 A11: DAC Heaters Not Regulated p. 285
2805 5700 Not Warmed Up p. 285
2806 A11: ADC Amp Output Noise p. 285
2807 A11: ADC Amp Output Offset p. 286
2808 A11: ADC Amp Gain Error p. 286
2809 A11: DAC Monitoring Fault p. 286
2810 A11: +11V DC Range Fault p. 286
2811 A11: -11V DC Range Fault p. 286
2812 A11: +22V DC Range Fault p. 286
2813 A11: -22V DC Range Fault p. 286
2814 A11: 6.5V Buffered Reference Fault p. 286
2815 A11: 6.5V Reference Fault p. 286
2816 A11: 13V Buffered Reference Fault p. 286
2817 A11: 13V Reference Fault p. 286
2818 Assembly A11 Missing p. 286
2819 A11: 8255 Was Reset p. 286
2820 A11: Fine Adjust Channel Fault p. 286
2821 A8/A11: +11/22V DC Zero Estimate Fault p. 286
2822 A8/A11: -11/22V DC Zero Estimate Fault p. 286
2823 A11: Could not Estimate +11V Or 22V Gain p. 286
2824 A11: Could not Estimate 6.5V Or 13V Ref p. 286
2825 A11: Could not Estimate 6.5V Or 13V Buf Ref p. 286
2826 A11: A/D Overload Fault p. 286
3101 A14: 8255 Port A Fault p. 286
3102 A14: 8255 Port B Fault p. 286
3103 A14: 8255 Port C Fault p. 286
3104 A15: HV Oven Regulation Fault p. 286
3105 A15: HV I Oven Regulation Fault p. 286
3106 A15: DC HV Amp Noise Fault p. 286
3107 A15: DC HV Amp Offset Fault p. 286
3108 A15: DC HV Amp Gain Fault p. 286
3109 A15: HV +DC Preamplifier Fault p. 286
3110 A15: HV +DC Series Pass & Current Fault p. 286
3111 A15: HV +DC High Voltage Output Fault p. 286
3112 A15: HV +DC Ref/Error Amplitude Fault p. 287
3113 A15: HV -DC Preamplifier Fault p. 287
3114 A15: HV -DC Series Pass & Current Fault p. 287
3115 A15: HV -DC High Voltage output Fault p. 287
3116 A15: HV -DC Reference/Error Amp Fault p. 287
3117 A14/A15: HV +DC Current Error Amp Fault p. 287
3118 A14/A15: HV -DC Current Error Amp Fault p. 287
3119 A14/A15: HV +DC Current Abs. Value p. 287
3120 A14/A15: HV -DC Current Abs. Value p. 287
3122 A14/A15: HV AC 1KHZ, Preamplifier Fault (lo) p. 287
3123 A14/A15: HV AC 1KHZ, Preamplifier Fault (mid) p. 287
3124 A14/A15: HV AC 1KHZ, Preamplifier Fault (hi) p. 287
3125 A14/A15: HV AC 100HZ, Preamplifier Fault (lo) p. 287
3126 A14/A15: HV AC 100HZ, Preamplifier Fault (hi) p. 287
3127 A14/A15: HV AC 1KHZ, Output Fault (lo) p. 287
3128 A14/A15: HV AC 1KHZ, Output Fault (mid) p. 287
3129 A14/A15: HV AC 1KHZ, Output Fault (hi) p. 287
3130 A14/A15: HV AC 100HZ, Output Fault (lo) p. 287
3131 A14/A15: HV AC 100HZ, Output Fault (mid) p. 287
3132 A14/A15: HV AC 100HZ, Output Fault (hi) p. 287
3135 Assembly A14 Not Responding p. 287
3136 A14: 8255 Was Reset p. 287
3137 A14/A15/A16: 2.2A AC Range Mag. Fault p. 287
3138 A14/A15/A16: 2.2A AC Range Compliance p. 287
3139 A14/A15/A16: 2.2A AC Range Amplitude p. 287
3140 A14/A15/A16: 2.2A AC Range Abs. Value p. 287
3141 A14/A15/A16: 2.2A DC Range Dummy Load p. 287
3142 A14/A15/A16: 2.2A DC Range Compliance p. 287
3143 A14/A15/A16: 2.2A DC Range Magnitude p. 287
3200 Assemblies Missing p. 288
3202 5725 ACV Sense Amp Fault p. 288
3203 5725 ACV Standby 5725 Fault p. 288
3204 5725 ACV Operate 5725 Fault p. 288
3205 5725 ACV Cal Sensor Test Died p. 288
3206 5725 ACV Cal Sensor Fault p. 288
3207 5725 ACV VLF (100 Hz) Won't Converge p. 288
3208 5725 ACV LF (1 kHz) Won't Converge p. 288
3209 5725 ACV MF (10 kHz) Won't Converge p. 288
3210 5725 ACV HF (100 kHz) Won't Converge p. 288
3211 5725 Current Path To Shunt Open p. 288
3212 5725 Shunt Sense Open p. 288
3213 5725 Shunt Measurement Out Of Tolerance p. 288
3214 5725 Current Amplifier Offset Too Large p. 288
3215 5725 Current Drive Path Open p. 288
3216 5725 Current Error Amplifier Failure p. 288
3217 5725 Amplifier Not Connected p. 288
3300 A9: 8255 Control Word p. 288
3301 A9: 8255 Port A Fault p. 288
3302 A9: 8255 Port B Fault p. 288
3303 A9: 8255 Port C Fault p. 288
3304 A9: 10V Source Fault p. 288
3305 A9: 5V Source Fault p. 288
3306 A9: 2V Source Fault p. 288
3308 A9: Diff Amp Offset Fault p. 288
3309 A9: Diff Amp Gain Fault p. 288
3310 A9: Diff Amp Noise Fault p. 288
3311 A9/A10: Ohms 10:1 Divider Fault p. 288
3312 A9/A10: Ohms 1:1 Divider Fault p. 288
3313 A9/A10: 10K Ohm Diagnostic Fault p. 289
3314 A9/A10: 19K Ohm Cal Diag Fault p. 289
3315 A9/A10: 10 Ohm Cal Diag Fault p. 289
3316 A9/A10: 19 Ohm Cal Diag Fault p. 289
3317 A9/A10: 100K Ohm Ratio Fault p. 289
3318 A9/A10: 190K Ohm Ratio Fault p. 289
3319 A9/A10: 1M Ohm Ratio Fault p. 289
3320 A9/A10: 1.9M Ohm Ratio Fault p. 289
3321 A9/A10: 10M Ohm Ratio Fault p. 289
3322 A9/A10: 19M Ohm Ratio Fault 1 p. 289
3323 A9/A10: 19M Ohm Ratio Fault 2 p. 289
3324 A9/A10: 100M Ohm Ratio Fault p. 289
3325 A9/A10: 10K Ohm Check Fault p. 289
3326 A9/A10: 19K Ohm Check Fault p. 289
3327 A9/A10: 1K Ohm Check Fault p. 289
3328 A9/A10: 1.9K Ohm Check Fault p. 289
3329 A9/A10: 100 Ohm Check Fault p. 289
3330 A9/A10: 190 Ohm Check Fault p. 289
3331 A9/A10: 10 Ohm Check Fault p. 289
3332 A9/A10: 19 Ohm Check Fault p. 289
3333 A9/A10: 1 Ohm Check Fault p. 289
3334 A9/A10: 1.9 Ohm Check Fault p. 289
3335 A9/A10: Ohms Short Check Fault p. 289
3336 A9/A10: 2 Wire Compensation Fault p. 289
3337 A9/A10: Ohms Correction Factor Fault p. 289
3338 Assembly A9 Not Responding p. 289
3339 A9: 8255 Was Reset p. 289
3340 A9/A10: 100 Ohm Cal Diag Fault p. 289
3341 A9/A10: 190 Ohm Cal Diag Fault p. 289
3400 A12: 8255 Control Word p. 290
3401 A12: 8255 Port A p. 290
3402 A12: 8255 Port B p. 290
3403 A12: 8255 Port C p. 290
3404 A13: 8255 Control Word p. 290
3405 A13: 8255 Port A p. 290
3406 A13: 8255 Port B p. 290
3407 A13: 8255 Port C p. 290
3408 A13: Fixed Ampl. Osc Fault p. 290
3409 A13: Phase Lock Loop Fault p. 290
3410 A12/A13: 22V Amp Bias Adj Error p. 290
3411 A12/A13: 22V Amp Nonfunctional p. 290
3412 A12 To A13 Interface Fault p. 290
3413 A12/A13: 14 bit DAC Nonfunctional p. 290
3414 A12/A13: Nonlinear Control Loop 2Vrng p. 290
3415 A12/A13: nonlinear Control Loop 20Vrng p. 290
3416 A12/A13: DAC 15th Bit Fault p. 290
3417 A12: DC Sensor Buffer Fault p. 290
3418 A12: Sensor Loop/Sq. Root Amp Fault p. 290
3419 A12: AC Sensor Buff (2V Range) p. 290
3420 A12: AC Sensor Buff (20V Range) p. 290
3421 A12: AC Cal Sensor (2V Range) p. 290
3422 A12: AC Cal Sensor (20V Range) p. 290
3423 Assembly A12 Missing p. 290
3424 Assembly A13 Missing p. 290
3425 Assembly A12 Or A13 Not Responding p. 290
3426 A12: 8255 Was Reset p. 290
3427 A13: 8255 Was Reset p. 290
3500 A16: 8255 Control Word p. 291
3501 A16: 8255 Port A p. 291
3502 A16: 8255 Port B p. 291
3503 A16: 8255 Port C p. 291
3507 A16: PA Supplies Are Off p. 291
3508 A16/A14: 220V AC Range Output Fault p. 291
3509 A16: Amplifier Loop Not Regulated p. 291
3510 A16: 220V Amp Fault p. 291
3511 A16: Incorrect PA Input p. 291
3520 A16: PA Oven Regulation Fault p. 291
3521 Assembly A16 Is Not Responding p. 291
3524 A16: Power Amp Is Too Hot p. 291
3525 220V DC Initialization Fault p. 291
3526 220V AC Initialization Fault p. 291
3527 A16: Power Amp DC Cal Network Zero Fault p. 291
3528 A16: Power Amp DC Cal Network Gain Fault p. 291
3529 A16: Pwr Amp 220V Range Attenuator Fault p. 291
3530 A16: 8255 Was Reset p. 291
3600 +17S Supply Fault p. 291
3601 -17S Supply Fault p. 291
3602 +15S Supply Fault p. 291
3603 -15S Supply Fault p. 291
3604 +42S Supply Fault p. 291
3605 -42S Supply Fault p. 291
3606 LH COM Ground Fault p. 291
3607 -5LH Supply Fault p. 291
3608 5RLH Supply Fault p. 291
3609 8RLH Supply Fault p. 291
3610 +PASupply Fault p. 291
3611 -PASupply Fault p. 291
3612 +15 OSC Supply Fault p. 291
3613 -15 OSC Supply Fault p. 292
3614 OSC COM Ground Fault p. 292
3615 S COM Ground Fault p. 292
3700 A21: 8255 Control Word p. 292
3701 A21: 8255 Port A Fault p. 292
3702 A21: 8255 Port B Fault p. 292
3703 A21: 8255 Port C Fault p. 292
3704 Assembly A21 Not Responding p. 292
3705 A21: Rear Panel Data Bus Fault p. 292
3708 A21: 8255 Was Reset p. 292
3800 A8: 8255 Control Word p. 292
3801 A8: 8255 Port A Fault p. 292
3802 A8: 8255 Port B Fault p. 292
3803 A8: 8255 Port C Fault p. 292
3804 A8: Zero Amp Lo Noise Fault p. 292
3805 A8: Zero Amp Lo Offset p. 292
3806 A8: Zero Amp Lo Gain Fault p. 292
3807 A8: Zero Amp Hi Noise Fault p. 292
3808 A8: Zero Amp Hi Offset p. 292
3809 A8: Zero Amp Hi Gain Fault p. 292
3810 A8: 2.2V Amp Noise Fault p. 292
3811 A8: Zero Amp Offset p. 292
3812 A8: 2.2V Gain Fault p. 292
3813 A8: 220mV Offset Fault p. 292
3814 A8: 220mV Divider Fault p. 292
3815 A8: 22mV Divider Fault p. 292
3818 A8: Out Lo To Sense Lo Continuity Fault p. 292
3819 A8: Relay Fault p. 292
3824 A8: Oven Regulation Fault p. 292
3825 Assembly A8 Not Responding p. 292
3826 A8: 8255 Was Reset p. 293
3827 Assembly A8 Too Hot p. 293
3900 A5: 8255 Control Word p. 293
3901 A5: 8255 Port A Fault p. 293
3902 A5: 8255 Port B Fault p. 293
3903 A5: 8255 Port C Fault p. 293
3904 Optional Assemblies A5/A6 Are Missing p. 293
3905 A6: Phase Lock Loop Fault p. 293
3906 A6: Phase Lock Loop Fault At 30MHz p. 293
3907 A5: RMS Sensor Fault p. 293
3908 A5: RMS Sensor Fault At 30MHz p. 293
3909 A5: RMS Sensor Fault At 6.5V DC in p. 293
3910 A5/A6: Ampl. Control Fault p. 293
3911 A5/A6: Ampl. Control Fault At 30MHz p. 293
3912 A5/A6: Output Offset Fault p. 293
3913 A5/A6: Output Offset Fault At 30MHz p. 293
3914 A5: 0DB Output Attenuation Fault p. 293
3915 A5: 10DB Output Attenuation Fault p. 293
3916 A5: 20DB Output Attenuation Fault p. 293
3917 A5: 30DB Output Attenuation Fault p. 293
3918 A5: 40DB Output Attenuation Fault p. 293
3922 A5/A6: Wideband Initialization Fault p. 293
3923 A5: 8255 Was Reset p. 293
4000 Bad Sequence ID p. 293
4001 Over Nested Subsequences p. 293
4002 Bad Sequence Command Code p. 293
4003 Bad Reply Size From Inguard p. 293
4004 Reply From Inguard Too Small p. 293
4005 False MSG Semaphore from Inguard p. 293
4006 Inguard CPU POP p. 293
4301 ROM Checksum p. 295
4302 RAM p. 295
4303 Real Time Clock p. 295
4304 Keyboard p. 295
4305 Output Display p. 295
4306 Control Display p. 295
4307 IEEE488 p. 295
4308 Rear Panel DUART p. 295
4309 Guard Crossing DUART p. 295
4310 Interrupt Hardware p. 295
4311 Inguard CPU ROM Fault p. 295
4312 Inguard CPU RAM Fault p. 295
4313 Front Panel Refresh Fault p. 295
4400 Invalid Command p. 295
4500 VFdcDac Value Out Of DAC Range p. 295
4501 Bad Range Selector p. 295
4502 Non-Vfinder Range p. 295
4503 Bad DAC ICALL p. 295
4504 User Range Adjust of 0 (I used 1) p. 295
4505 Divide By 0 In (VF)correct p. 295
4600 Watchdog Timed Out p. 295
4700 Kstart Returned p. 295
4701 Task Termination p. 295
4702 No Signal Handler p. 295
4703 Ksuspend( ) When Not RUNNING p. 295
4704 Ksleep( ) When Not RUNNING p. 295
4705 Kewait( ) When Not RUNNING p. 295

Error codes from p. 284

Codes and descriptions are taken from the manufacturer’s manual; page numbers link to the PDF. ManualLib is not affiliated with Fluke. Found a mistake? Report an error.