Bio‑Rad 164-5050 Error E9

Electrophoresis power supply · taken from the manufacturer’s manual, pages cited

Error codeE9

Change in Load Resistance indicates a potential failure of the cell's power leads or a loose output connection. The manual's first action is: “Check and correct any potential resistance problem then,
• press run/pause key to resume run or,
• press any key, other than the run/pause key, to clear the error code.”

Model
Bio‑Rad 164-5050
Display shows
E9
Type
Error code
Source
Manual p. 13

What E9 means

Change in Load Resistance.
The PowerPac Basic detects drastic changes in resistance which may indicate failure of the cell's power leads or a loose output connection.
Note: Certain applications exhibit fluctuations in resistance that can trigger the change in resistance error code. If this is the case, the change in resistance feature can be deactivated to allow uninterrupted completion of a run. (see Section 3, Step 8)
Caution: Deactivating this safety feature increases the possibility that a failure of the power leads or loose output connection will not be detected. p. 13

Common causes

  • Change in Load Resistance.
    The PowerPac Basic detects drastic changes in resistance which may indicate failure of the cell's power leads or a loose output connection.
    Note: Certain applications exhibit fluctuations in resistance that can trigger the change in resistance error code. If this is the case, the change in resistance feature can be deactivated to allow uninterrupted completion of a run. (see Section 3, Step 8)
    Caution: Deactivating this safety feature increases the possibility that a failure of the power leads or loose output connection will not be detected. p. 13

What to do

  1. Check and correct any potential resistance problem then,
    • press run/pause key to resume run or,
    • press any key, other than the run/pause key, to clear the error code.
    p. 13

If E9 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio‑Rad service.

This page summarises what the manufacturer’s manual says about E9; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio‑Rad. Found a mistake? Report an error.