Internal Short Circuit indicates a short circuit within the system, possibly due to dirty contacts or frayed wires. The manual's first action is: “Check for and correct problems such as dirty contacts, frayed wires, excessive buffer concentration. Then press any key other than the run/pause key to clear the code.”
- Model
- Bio‑Rad 164-5050
- Display shows
- Internal Short Circuit (E13)
- Type
- Error code
- Source
- Manual p. 13
What E13 means
Internal Short Circuit p. 13
Common causes
- Internal Short Circuit p. 13
What to do
- Check for and correct problems such as dirty contacts, frayed wires, excessive buffer concentration. Then press any key other than the run/pause key to clear the code. p. 13
If E13 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio‑Rad service.
This page summarises what the manufacturer’s manual says about E13; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio‑Rad. Found a mistake? Report an error.