Uni-Trend UT513 Insulation Resistance Tester Operating Manual
This manual provides instructions for operating the Uni-Trend UT513 insulation resistance tester. It includes sections on unpacking, safety precautions, meter structure, display, key functions, measurement operations, USB interface, and maintenance. Technical specifications for voltage and insulation resistance measurements are also detailed.
The device is designed for handheld insulation resistance testing and complies with IEC61010 safety standards. It features battery operation with sleep mode, USB connectivity, and a backlit LCD display. Safety warnings and operational cautions are emphasized to prevent electric shock and ensure accurate measurements.
Uni-Trend UT513 specifications
| Safety Compliance | EN 61010-1:2010, Pollution Degree 2, CAT III 600V, double insulation |
|---|---|
| Battery | 8 x 1.5V LR14 batteries |
| Power Adaptor Input | 230V, 50/60Hz, 75mA |
| Power Adaptor Output | DC 13.5V, 1100mA |
| Output Voltages | 500V, 1000V, 2500V, 5000V |
| Insulation Resistance Range at 5000V | 10M ~ 1000G, open circuit voltage DC 5000V |
| Current Consumption | Maximum around 1.0A, average around 20mA |
Source: Uni-Trend UT513 Insulation Resistance Tester Operating Manual
All Uni-Trend UT513 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What items are included with the UT513?
The package includes the English operating manual, three test leads with alligator clips (black, green and red), eight 1.5V LR14 batteries, a tool box, a USB interface cable, software and optionally a power adaptor.
From: Uni-Trend UT513 Insulation Resistance Tester Operating Manual
What safety category and insulation does the meter comply with?
It complies with EN 61010-1:2010 for pollution degree 2, measurement category CAT III 600V and double insulation.
From: Uni-Trend UT513 Insulation Resistance Tester Operating Manual
What is the battery indicator status when the battery is empty?
The battery indicator shows 10V or less, meaning the battery is empty and the meter cannot guarantee accuracy.
From: Uni-Trend UT513 Insulation Resistance Tester Operating Manual
What safety standard does the UT513 comply with?
The UT513 complies with IEC 61010 safety measurement requirement: in pollution degree 2, overvoltage category (CAT. III 600V) and double insulation.
From: Uni-Trend UT513 Insulation Resistance Tester Operating Manual