Agilent GTA-96 Graphite Tube Atomizer Application Note for Trace Metals in Palm Oil
The document describes a method for monitoring metal contamination in various types of palm oil (crude, refined, degummed, hydrogenated) caused by processing steps like bleaching and deodorization. The analysis utilizes Graphite Furnace Atomic Absorption Spectroscopy (GFAAS), which offers advantages such as requiring small sample quantities and providing greater sensitivity compared to flame methods.
The note outlines specific instrumentation used, including the Agilent SpectrAA-40 spectrophotometer with hollow cathode lamps and an Epson MX-80 printer. It details sample preparation techniques involving dissolution in solvents like DIBK, MIBK, or xylene, and provides optimal furnace program parameters, specifically ashing temperatures (700°C for Iron/Copper, 800°C for Nickel), to ensure accurate results by minimizing background signals while preventing analyte loss.
The GTA-96 specifications
| Spectrometer Model | Agilent SpectrAA-40 |
|---|---|
| Silicon Flame Wavelength | 251.6 nm |
| Tin Furnace Wavelength | 251.6 nm |
| Titanium Furnace Wavelength | 364.3 nm |
Source: Agilent GTA-96 Silicon Tin Titanium Jet-Engine Oil Analysis Application Note
All The GTA-96 documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What volume of sample solution should be dispensed for analysis?
The total dispensed volume should be less than 15 pL to avoid requiring very long drying and ashing times.
From: Agilent GTA-96 Graphite Tube Atomizer Application Note for Trace Metals in Palm Oil
What is the recommended dilution ratio for palm oil samples?
A suitable dilution is 1 g of sample per 5 mL of solution to balance concentration and precipitation risks.
From: Agilent GTA-96 Graphite Tube Atomizer Application Note for Trace Metals in Palm Oil
Which elements are analyzed in this application note?
The document focuses on the determination of silicon, tin, and titanium in jet-engine oils.
From: Agilent GTA-96 Silicon Tin Titanium Jet-Engine Oil Analysis Application Note
What is the recommended sample volume for organic solvents?
Small sample volumes between 10 and 20 microlitres are recommended to prevent solvent creep errors.
From: Agilent GTA-96 Silicon Tin Titanium Jet-Engine Oil Analysis Application Note