Tangshan Dafang Huizhong LZD127 Mining Explosion-Proof IS Flow Meter Installation and Operating Instructions
The LZD127 is a mining explosion-proof and intrinsically safe electromagnetic flow meter made by Tangshan Dafang Huizhong Instrument Co., Ltd. It measures the flow of conductive liquids in circular pipes using Faraday's law of electromagnetic induction and can be installed in coal mine underground and other explosive hazardous locations. The manual covers the working principle, operating environment, technical specifications, model selection, converter and sensor installation, electrical connection, converter operation, maintenance and fault diagnosis, and the RS485 communication protocol.
The document lists models from LZD127/25G to LZD127/300G for pipe sizes from 25 mm to 300 mm. It states a flow velocity range of 0.3 m/s to 10.0 m/s, 0.5-class accuracy, a 127 V AC 50 Hz power supply with no more than 10 W consumption, IP54 protection, and Exd[ib]IMb explosion protection. It also gives installation precautions, straight-pipe requirements, terminal wiring descriptions, and safety warnings such as not opening the instrument while it is energized. This manual is written in Chinese.
Tangshan Dafang Huizhong Instrument LZD127 specifications
| Instrument name | Mining explosion-proof and intrinsically safe flow meter |
|---|---|
| Models | LZD127/25G, LZD127/50G, LZD127/80G, LZD127/100G, LZD127/150G, LZD127/200G, LZD127/250G, LZD127/300G |
| Pipe diameters | 25, 50, 80, 100, 150, 200, 250, 300 mm |
| Flow velocity range | 0.3 m/s to 10.0 m/s |
| Accuracy | 0.5 class (±0.5% indicated error) |
| Power supply | 127 V AC, 50 Hz; ≤10 W |
| Protection rating | IP54 |
| Explosion protection marking | Exd[ib]IMb |
| Minimum medium conductivity | 5 μs/cm |
| Data retention after power loss | ≥10 years |
All Tangshan Dafang Huizhong Instrument LZD127 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What liquids can the LZD127 flow meter measure?
It measures conductive liquids with conductivity of at least 5 μs/cm, such as acid, alkali, salt solutions, slurry, wastewater, and tap water. Pure alcohol, acetone, and oils cannot be measured.
Where can the LZD127 be installed?
It can be installed in coal mine underground and other locations with explosive hazards, within the stated environmental conditions.
What is the minimum conductivity required for the measurement medium?
The conductivity of the medium cannot be lower than 5μs/cm.
From: Tangshan Dafang Huizhong LZD127 Explosion-proof Flowmeter Installation and Operating Instructions
Can the flowmeter be used in environments with strong magnetic fields?
No, it should be installed in an environment without strong magnetic field interference.
From: Tangshan Dafang Huizhong LZD127 Explosion-proof Flowmeter Installation and Operating Instructions