LZD127 mine-type explosion-proof and intrinsically safe flow meter is a magnetic flow meter for measuring the flow of liquid in circular pipes on the principle of Faraday's law of electromagnetic induction. Its entire design meets GB3836.1-2010, GB3836.2-2010 and GB3836.4-2010 requirements for explosive environments. It can be installed and used in coal mine pits and other places with explosive hazards.
The LZD127 is a mining explosion-proof and intrinsically safe electromagnetic flow meter made by Tangshan Dafang Huizhong Instrument Co., Ltd. It measures the flow of conductive liquids in circular pipes using Faraday's law of electromagnetic induction and can be installed in coal mine underground and other explosive hazardous locations. The manual covers the working principle, operating environment, technical specifications, model selection, converter and sensor installation, electrical connection, converter operation, maintenance and fault diagnosis, and the RS485 communication protocol.
The document lists models from LZD127/25G to LZD127/300G for pipe sizes from 25 mm to 300 mm. It states a flow velocity range of 0.3 m/s to 10.0 m/s, 0.5-class accuracy, a 127 V AC 50 Hz power supply with no more than 10 W consumption, IP54 protection, and Exd[ib]IMb explosion protection. It also gives installation precautions, straight-pipe requirements, terminal wiring descriptions, and safety warnings such as not opening the instrument while it is energized. This manual is written in Chinese.
| Instrument name | Mining explosion-proof and intrinsically safe flow meter |
|---|---|
| Models | LZD127/25G, LZD127/50G, LZD127/80G, LZD127/100G, LZD127/150G, LZD127/200G, LZD127/250G, LZD127/300G |
| Pipe diameters | 25, 50, 80, 100, 150, 200, 250, 300 mm |
| Flow velocity range | 0.3 m/s to 10.0 m/s |
| Accuracy | 0.5 class (±0.5% indicated error) |
| Power supply | 127 V AC, 50 Hz; ≤10 W |
| Protection rating | IP54 |
| Explosion protection marking | Exd[ib]IMb |
| Minimum medium conductivity | 5 μs/cm |
| Data retention after power loss | ≥10 years |
It measures conductive liquids with conductivity of at least 5 μs/cm, such as acid, alkali, salt solutions, slurry, wastewater, and tap water. Pure alcohol, acetone, and oils cannot be measured.
It can be installed in coal mine underground and other locations with explosive hazards, within the stated environmental conditions.
The upstream straight pipe should be at least 5DN; if a non-full-open gate or control valve is upstream, increase it to 10DN. The downstream straight pipe should generally be at least 3DN.
It requires 127 V AC, 50 Hz, with power consumption no more than 10 W.