Key facts
- Brand
- Runze
- Model
- Smart SY-01
- Category
- Other products
- Documents
- 2
- Languages
- ZH
- Last added
- 2023-11-07
Start here

Runze Smart SY-01B Syringe Pump Quick Start Guide
This quick start guide for the Runze Smart SY-01B syringe pump lists the delivered items, explains how the power and communication cables are wired, and walks through installing the USB-to-RS232/RS485 driver and the serialcomm debugging software from the runzeliuti.com website. It then shows example serial commands, such as the /1ZR reset command and commands for switching the valve channel and moving the piston, along with a table of quick commands.
The guide also gives basic troubleshooting for a pump that does not power on, for commands that get no reply, and for RS232/RS485 wiring mistakes, and closes with precautions. Among them: use a 24 V DC ±10%, 3 A supply, keep the ground wire connected, do not hot-plug the DB15 connector, and flush the valve channels and syringe with water or alcohol after use. This manual is written in Chinese.
Runze Smart SY-01 specifications
| Liquid volume accuracy | Error ≤1% (rated stroke) |
|---|---|
| Liquid volume precision (repeatability error) | 0.3%-0.5% |
| Rated stroke (control steps) | 30 mm (12000 steps) |
| Linear speed | 0.017 mm/s-4.167 mm/s (pure water) |
| Rated stroke run time | 7 s-1765 s (pure water) |
| Control resolution / minimum injection precision | 0.0025 mm (1 step) |
| Drive mechanism | Ball screw (lead 1 mm) |
| Valve switching time | ≤280 ms (adjacent ports) |
| Maximum liquid path pressure | 0.7 MPa (reference) |
| Wetted materials | Borosilicate glass, PCTFE, sapphire |
| Valve head types (optional) | M01, M02, M03, M04, M05, M06, M10 |
| Compatible syringes (optional) | 25 µl to 5 ml imported syringes |
| Tube interface | 1/4-28UNF internal thread |
| Syringe interface | 1/4-28UNF internal thread |
| Signal input | 3 TTL signal inputs |
| Signal output | 3 TTL signal outputs |
| Communication interfaces | RS232/RS485/CAN bus |
| Power supply | DC24V/3A |
| Operating temperature | 5°C-55°C |
| Operating relative humidity | <80% |
| Dimensions (L*W*H) | 137*45*127 mm |
| Weight | 1.13 kg |
All Runze Smart SY-01 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
The shortest path to a working setup, usually only a few pages.
Questions answered in these manuals
What is the default communication address and baud rate?
The default device address is 00 and the baud rate is 9600.
What power supply does the pump require?
It requires a DC24V/3A power supply.
What is the factory default baud rate of the pump?
The factory default baud rate is 9600 bps; it is set in the debugging software together with the serial port.
What power supply does the SY-01B require?
It requires a 24 V DC ±10%, 3 A supply, and when a linear power supply is used the voltage and current must be set to those values.