Ocean Optics USB4000 Manuals

ManualLib has 2 documents for the Ocean Optics USB4000 (Analyzers): 1 datasheet, 1 user manual. Every file can be read online or downloaded as a PDF for free.

Datasheet 1User Manual 1

Key facts

Model
USB4000
Category
Analyzers
Documents
2
Languages
EN
Last added
2026-10-03

Start here

Ocean Optics USB4000 Fiber Optic Spectrometer Installation and Operation Manual

Ocean Optics USB4000 Fiber Optic Spectrometer Installation and Operation Manual

User Manual 36 pages · 851 KB · EN

This manual provides instructions for setting up, calibrating and performing experiments with the Ocean Optics USB4000 Spectrometer, a miniature fiber-optic spectrometer with a 3648-element TCD1304AP detector, high-speed electronics and USB or serial port connectivity to a computer.

The document describes the product overview and features, system requirements and interface options, installation steps for USB and serial port modes, troubleshooting for Windows, Mac and Linux systems, wavelength calibration procedures, external triggering modes, and technical specifications including the 22-pin accessory connector pinout.

Read online

Ocean Optics USB4000 specifications

Wavelength range200 nm – 1100 nm
Resolution (FWHM)~0.3 – 10.0 nm
Power consumption250 mA @ 5 VDC
Integration time10 µs – 65 s
A/D converter16‑bit, 3 MHz
Programmable GPIO pins6 user‑programmable digital I/O pins

Source: Ocean Optics USB4000 Spectrometer Data Sheet

All Ocean Optics USB4000 documents by type

Datasheet 1

Ratings, dimensions and ordering codes, for selection and design work.

User Manual 1

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What power source does the USB4000 use?

It operates from a single +5 VDC supply provided via the USB connection or from an external 5‑V power source.

From: Ocean Optics USB4000 Spectrometer Data Sheet

How many programmable digital I/O pins are available?

The spectrometer provides six user‑programmable digital input/output pins.

From: Ocean Optics USB4000 Spectrometer Data Sheet

What operating systems does the USB4000 support?

Windows 98/Me/2000/XP, Apple Macintosh OS X 10.0 or later, and Linux distributions including Red Hat 9 or later, Fedora, Debian 3.1 (Sarge) and SUSE 9.0 or later, with Ocean Optics' SpectraSuite software installed.

From: Ocean Optics USB4000 Fiber Optic Spectrometer Installation and Operation Manual

Does the USB4000 need an external power supply?

No. When connected through USB 2.0 or 1.1, the spectrometer draws power from the host computer, eliminating the need for an external power supply.

From: Ocean Optics USB4000 Fiber Optic Spectrometer Installation and Operation Manual